28
MIDEX
Kirchgaesser Industrieelektronik GmbH
Controls
4.5.9 Function 3-9: Pulse length
in-/outputs
3-9
pulse length
Adjustment of the pulse length
Code number to display function:
none
Code number to modify parameters: user
Available range: 0..99.999.999 ms in steps of 10 ms.
Example:
pulse length
1000 ms
Pulse length = 1 second
The pulse length is limited through the pulse unit and
the expected maximum flow rate.
4.5.10 Function 3-10: Volume counter (only MIDEX-**T**)
in-/outputs
3-10
volume counter
Adjustment of the start value for the volume counter.
Code number to display function:
user
Code number to modify parameters: user
Available range: - 1.999.999.999 .. + 1.999.999.999
Example:
volume counter
+10000
Adjust the start value of the volume counter to 10.000
(the unit is dependent on the pulse unit)
4.5.11 Function 3-12: Default in-/outputs
in-/outputs
3-12
load defaults
Code number to display function:
user
Code number to modify parameters: user
After safety query the default parameters for in-/out-
puts are loaded from the EPROM to the EEPROM and
the RAM.
You find an overview of the default values in the section 4.9 of this chapter.
Are you sure?
<E> = yes