Kipp & Zonen DustIQ Instruction Sheet Download Page 2

2

/4

www.kippzonen.com/dustiq

Needed before tare

1. Connected, working clean DustIQ with software version 2.15 or higher

2. DustIQ must be operative for minimal 3 minutes

3. Modbus registers 20-40 must be read and logged for soiling ratio’s, transmission losses, status flags,

   dust slopes etc. 

(see manual section 4.2)

4. A person on-site        and a remote person reading DustIQ data and the manual       .

5. If moving from the back of the DustIQ to the front takes more than 20 seconds, then consider adding  

  a third person, just to push the button.

DustIQ LED light interpretation

Regular operation, 5 seconds of blue led light with a 60 second interval

Feedback during the Tare or field calibration are given by the  2 sensors and the calibration button. 

There are 2 cases:

1. Slow blinking is 1x per second, confirmation and DustIQ awaiting next input

2. Fast blinking is 10x per second, this is the error signal 

  for further information see error code in register 27 

Both signals last 30 seconds in total, before normal operation restarts.

1

2

12 1

5

7

11

2

4

8

10

6

3

9

15 min

Time needed is 15 minutes

Die benötigte Zeit beträgt 15 min

Le temps nécessaire est de 15 min

El tiempo necesario es de 15 min

所需时间为

15

分钟

Clear, sunny day

Klarer, sonniger Tag

Journée claire et ensoleillée

Día claro y soleado

晴朗的晴天

No clouds

Keine Wolken

Pas de nuages

Sin nubes

无云

Between 2 hours before and after solar noon

Zwischen 2 Stunden vor und nach Sonnenmittag

Entre 2 heures avant et après midi solaire

Entre 2 horas antes y después del mediodía solar.

中午前后2小时之间

Clean, dry cloth

Sauberes, trockenes Tuch

Chiffon propre et sec

Paño limpio y seco

用干布擦拭清洁

Clean, demineralised water and a sponge

Sauberes, demineralisiertes Wasser und ein Schwamm

De l'eau propre et déminéralisée et une éponge

Agua limpia y desmineralizada y una esponja.

用海绵和淡水清洁

12 1

11

2

3

Example

Solar noon: 1 PM

Validate installation

No action needed

Wait

Wait 3 full minutes for the DustIQ to stabilize.

Wait

Sensors and button light up continuously for 1 minute.

Push the Calibration button on the underside of the DustIQ 3 times 
in rapid succession.

No action needed

Push button 3 times

No action needed

No action needed

Check sensors or button

Slow blinking = OK and finished.
Inform person        of successful completion.

Fast blinking = error.
Inform person       .

Confirm successful calibration in register 27 (manual section 4.2.2.)

If unsuccessful read error code in register 27 for possible reasons and solutions.

On-Site

1

2

2

Remote office (or direct connection)

2

Sensor 1

Sensor 2

TO

P

Sensor 2

Host

Daisy chain -or-

Temperature sensor

Sensor 1

Solar panel

Calibration button

3-5 sec

Blue LED ring follows blue sensor LEDs

Calibration

During the procedure:

- Do not cast a shadow on the DustIQ

- Do not cover the sensors

1. The DustIQ is mounted properly as per instructions.
2. Power cable is connected to the DustIQ.
3. DustIQ is cleaned.

Do NOT block the sunlight on the DustIQ

TARE INSTRUCTIONS

Reviews: