KINLEE FT3010 User Manual Download Page 6

-11- 

8.2.3 Offset value setting---Menu F3

Default 0.0

 (0.0

)

 

Enter into setting mode, press “SET” key triple, the LCD display “ F3”, 

then press “+” key to increase the setting value 0.1

(0.1

), press “-” 

key to reduce the setting value 0.1

(0.1

). 

8.2.4 Sound notification setting 

Enter into setting mode, press “SET” key four times, the LCD display 

“  F4”,  then  press  “+”  key  to  turn  on  the  sound  notification  function, 

press “-” key to turn off the sound notification function. 

8.2.5 Backlight setting---Menu F5 

Enter into setting  mode, press “SET” key five times, the LCD display 

“ F5”, then press “+” key to turn on the backlight function, press “-” key 

to turn off the backlight function. 

8.2.6 Exit setting mode 

Press “SET” key until shutdown. 

 

8.3 Advice For The 1

st

 Use 

For  stable  and  reliable  results,  it  is  essential  to  check  the  forehead 

thermometer and change as needed, as follows: 

8.3.1

  Take  the  temperature  of  the  same  person  using  a  conventional 

thermometer; you will get 37.0

℃(

98.6

℉)

for instance

8.3.2

  Take  the  temperature  of  the  same  person  using  the  forehead 

thermometer  keeping  the  50  to  150  mm  distance  between  the 

thermometer  and  the  forehead  (Take  care  to  remove  any  obstacle  which 

could alter the measurement (hair, perspiration…). If you get 37.0

℃(

98.6

℉)

,the Non-contact body infrared Thermometer is properly set and ready 

for use. 

  If  you  get  a  lower  temperature,  36.4

℃(

97.4

℉)

for  example,  your 

difference  is  0.6

℃(

1.2

℉)

,  you  should  adjust  the  temperature  on  the 

forehead Thermometer Body infrared Thermometer and add the difference, 

i.e. 0.6

℃(

1.2

℉)

-12- 

To  do  it,  press  the  “SET”  button  for  3  seconds,  the  screen  displays  F1, 

press “SET” button again until you get F3, press “+” button in order to add 

the difference (in our example, 0.6

℃(

1.2

℉)

8.3.3  To  check,  take  the  temperature  again  using  the  Non-contact  body 

infrared Thermometer. 

 

Caution

This setting applies only to Body mode

 

8.4 Correct measurement 

8.4.1 Measurement body temperature   

Hold  the  forehead  thermometer,  leave  one  finger  in  front  of  the  button, 

keep  transducer  window  at  human’s  forehead,  leaving  a  distance  of 

50-150mm from the  forehead. Press the button, and the thermometer will 

start up automatically and display the measurement result, as shown in the 

figure below:   

 

 

 

 

                                                             

 

8.4.2 Measurement ambient temperature. 

Hold  the  thermometer,  leave  one  finger  in  front  of  the  button,  Level  the 

detection window face to the place where you want to measure,  and then 

press  button. The thermometer will start up  automatically and display the 

measurement result, as shown in the figure below:   

 

 

 

 

 

Summary of Contents for FT3010

Page 1: ...acteristics 6 5 Technical Data 7 6 Intended Use 8 7 Parts Identification As picture 9 7 1 Basic components 9 7 2 Display Indicator 10 8 Operation and setting method 10 8 1 Installing or replacing the batteries 10 8 2 Setting method 10 8 3 Advice For The 1st Use 11 8 4 Correct measurement 12 9 Troubleshooting 14 10 Care and Maintenance 14 11 Quality promised and service 15 12 Attachment list 15 13 ...

Page 2: ...ated in accordance with EN60601 1 2 2007 for EMC This does not guarantee in any way that the device will not be affected by electromagnetic interference Avoid using the device in high electromagnetic environment 3 Important Safety Information The warning symbol and cutline mentions in this user manual give guidance of safety and correct method and avoid harm to your body The warning symbol and cut...

Page 3: ...nit Do not disassemble repair or modify the unit Please use and keep this product with the standard mentions in this manual Otherwise It can not measure correctly 6 Meanings of Symbols Signs Notes on the signs Type B Applied Part Operating instructions Manufacturer Prohibition of free throw European community Specific batch code CE0120 CE Mark 4 Product Characteristics Designed for measuring the f...

Page 4: ...urface temperature Forehead temperature or object surface temperature 8 WARNING the performance of the instrument may be adversely affected the following occur Operation outside of the specified subject temperature t Operation outside of the specified ambient temperature and humidity ranges Storage outside of the specified ambient temperature and humidity ranges Mechanical shock may be adversely a...

Page 5: ... from our company 7 Parts Identification As picture 7 1 Basic components 10 7 2 Display 8 Operation and setting method 8 1 Installing or replacing the batteries a Push the battery cover back open the cover from the body b Insert two AA batteries as indicated in the battery compartment and then replace the battery cover 8 2 Setting method Keep pressing SET key and hold for 3 seconds into the settin...

Page 6: ...le which could alter the measurement hair perspiration If you get 37 0 98 6 the Non contact body infrared Thermometer is properly set and ready for use If you get a lower temperature 36 4 97 4 for example your difference is 0 6 1 2 you should adjust the temperature on the forehead Thermometer Body infrared Thermometer and add the difference i e 0 6 1 2 12 To do it press the SET button for 3 second...

Page 7: ...ou can measure temperature more accurately If the measurement is made without an effective distance or an deflection of the central position of the measured object a deflection from actual temperature may cause It is suggested to repeat the measurement once or more While measuring it is suggested to focus the detection window at the forehead and press the measurement button in front of the handle ...

Page 8: ...n by the non formal used or disassemble by you the device can not be in the guarantee range Tips Please keep the guarantee card and the buying receipt which is very useful when you need to amend it 12 Attachment list One set of Electronic Forehead Infrared Thermometer One set of user manual One set of the Warranty certificate 2 1 5 AA Batteries optional 16 13 Safety classification of ME EQUIPMENT ...

Reviews: