![KINLEE FT3010 User Manual Download Page 5](http://html1.mh-extra.com/html/kinlee/ft3010/ft3010_user-manual_1971313005.webp)
-9-
Normal Temperatures According To Age
Age
Temp
℃
Temp
℉
0-2 years
36.4 to 38.0
97.5 to 100.4
3-10 years
36.1 to 37.8
97.0 to 100.0
11-65 years
35.9 to 37.6
96.6 to 99.7
﹥
65 years
35.8 to 37.5
96.4 to 99.5
Attention:
ASTM laboratory accuracy requirements in the display of 37
to 39
℃
(98 to 102
℉
) for Skin IR thermometers is
±
0.3
℃
(±
0.5
℉
),
whereas for mercury in-glass and electronic thermometers, the
requirement per ASTM Standards E667-86 and 1112-86 is
±
0.1
℃
(±
0.2
℉
)
.
When you need to clinical accuracy characteristics and procedures,
please contact with the local distributors obtain from our company.
7. Parts Identification (As picture)
7.1
Basic components
-10-
7.2
Display
8. Operation and setting method
8.1
Installing or replacing the
batteries
a. Push the battery cover back,
open the cover from the body.
b. Insert two “AA” batteries as indicated
in the battery compartment and then replace
the battery cover.
8.2 Setting method
Keep pressing “SET” key and hold for 3
seconds into the setting mode.
8.2.1 Select measurement unit---Menu F1
Enter into setting mode, the LCD displays “F1”, press “+” key to select
“
℉
” unit, press “-” key to select “
℃
” unit.
8.2.2 Warning value setting---Menu F2
【
Default 37.8
℃
(100.4
℉
)
】
Enter into setting mode, short press “SET” key twice, the LCD
displays “ F2”, then press “+” key to increase the setting value
0.1
℃
(0.1
℉
), press “-” key to reduce the setting value 0.1
℃
(0.1
℉
).