Notes
1.
Fan self-detect
When the system is turned on, Fan1 and Fan2 will start for about 10
seconds to check themselves (the fan will bring out the airflow), and
then stop automatically. (This self-test will not show the fan
operation message on the display panel). If the fans keep
not-working, please contact the after-sales for technical support.
2.
Fans work only when needed
System will detect the power loading automatically. If the
temperatures get high, system will active a fan for cooling until the
power loading getting low. Once the system detects that the
activated fan has stopped, a notice will pop up on any screen to show
Fan 1 Stop, Fan 2 Stop, or Fan 1/2 Stop on any screen as shown in the
images below:
The setting operation can still be used normally and will not be
affected by these message.
3.
Fan1 Stop:
The location of Fan 1 is as shown in the yellow frame:
Fan 1 is located at the top of the trolley and can be seen from the
exterior. If “Fan 1 Stop” is displayed, please restart the system and