background image

INTRODUCTION

INTRODUCTION

7

6

CAUTION

* If you experience discomfort during a measurement, such as pain in the arm or other complaints, press 

the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from 

your arm.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not 

deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the arm and press the 

START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. Check the device, 

do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause 

injury, improper results, or serious danger.

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The 

typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after maintenance and 

repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air 

leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local 

guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, descriptions, 

calibration instructions,etc., to assist to service personnel in parts repair.

* The operator shall not touch output of batteries and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean 

the whole unit before and after use. Don’t use any abrasive or volatile cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please contact the 

SERVICE PERSONNEL of Harvard Medical Devices. Don’t open or repair the device by yourself in the 

event of malfunctions. The device must only be serviced, repaired and opened by individuals at

authorized sales/service centers. 

* Please report to Harvard Medical Devices if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small 

parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage temperature between 

uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the 

maximum storage temperature between uses until it is ready for intended use.

* This equipment needs to be installed and put into service in accordance with the information provided in 

the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile phones, cordless 

telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a 

distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80 

MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE. 

Otherwise, it may cause damage to the unit or danger to the user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a possibility that they 

might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood 

vessel.

* Please use the device under the environment which was provided in the user manual. Otherwise, the 

performance and lifetime of the device will be impacted and reduced.

SYMBOL

DESCRIPTION

Systolic pressure

High blood pressure 

Diastolic pressure

Low blood pressure

EXPLANATION

mmHg

Measurement Unit of the blood pressure

Low battery

Batteries are low and need to be replaced

Current Time

Year/Month/Day, Hour : Minute

(1mmHg=0.133kPa)

Heartbeat

Irregular heartbeat 

Average value

The average value of blood pressure

Pulse in beats per minute

Pulse display

Blood pressure monitor is detecting an 
irregular heartbeat during measurement.

Blood pressure 
level indicator

Indicate the blood pressure level

Blood pressure monitor is detecting a 
heartbeat during measurement.

mmHg

mmHg

/min

LCD display signal

Summary of Contents for WBP1

Page 1: ...Blood Pressure Monitor Instruction Manual WBP1...

Page 2: ...MENT 15 16 Check the memory Delete the Readings INFORMATION FOR USER 17 Tips for measurement Maintenance ABOUT BLOOD PRESSURE 18 19 What are systolic pressure and diastolic pressure What is the standa...

Page 3: ...in measuring blood pressure and heartbeat rate with arm circumference 22cm to 42cm about 8 16 It is intended for adult indoor use only 1 The device is not suitable for use on may be pregnant women or...

Page 4: ...st be treated with caution Observe the storage and operating conditions described in this booklet The equipment is not AP APG equipment and not suitable for use in the presence of a flammable anesthet...

Page 5: ...d sales service centers Please report to Harvard Medical Devices if any unexpected operation or events occur Keep the unit out of reach of infants young children or pets to avoid inhalation or swallow...

Page 6: ...is dim The display does not light up 1 Battery powered mode 6VDC 4 AAA batteries Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batterie...

Page 7: ...is stored in the memory The setting range of the year 2017 2057 Time format 24H the YEAR Each press will increase the numeral by one in a cycling manner mmHg mmHg min for 3 seconds to enter the mode...

Page 8: ...same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurements This allows you...

Page 9: ...off otherwise it will turn off within 1 minute 1 When the monitor is off please press button to show the average value of the latest three records If the records are less than 3 groups it will displa...

Page 10: ...ken in the following circumstances 1 Within 1 hour after drinking or eating 2 Immediate measurement after tea coffee or smoking 3 Within 20 minutes after taking a bath 4 When talking or moving your ng...

Page 11: ...records all the pulse intervals and calculate the average if there are two or more pulse intervals the difference between each interval and the average is more than the average value of 25 or there a...

Page 12: ...nd frequently asked questions for problems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing Type BF a...

Page 13: ...cation of usability engineering to medical devices EN 62304 2006 AC 2008 IEC 62304 2006 A1 2015 Medical device software Software life cycle processes Risk management Labeling User manual General Requi...

Page 14: ...lt Technical description Table 1 Group I Class B Not application Not application 1 all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFOR MANCE with regard to electromagnetic dis...

Page 15: ...2 0 3 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse m...

Reviews: