5
•
Reorient or relocate the receiving device.
•
Increase the separation between the equipment.
•
Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.
•
Consult the manufacturer for help.
The use of portable and mobile RF equipment may have an impact on this and other pieces of medical equipment.
The use of accessories, cables, or transducers other than those specified in this manual can significantly increase
emissions performance and degrade immunity performance of the product. Also, by using an accessory, transducer,
or cable with the product, other than those specified in this manual, it becomes the responsibility of the third-party
supplier or the user of the product, to determine compliance with the requirements of IEC 60601-1-2 when using this
item.
CAUTION
It is prudent to separate all electrical equipment that is very close in distance to the QTUG system. If it is essential to use the QTUG
system very close to other electrical equipment, it is prudent to determine, by observation, if the performance of either product is
affected by unintended electromagnetic coupling.
WARNING
It is advised not to use equipment other than the following devices listed by manufacturer (Kinesis QTUG sensors with Bluetooth
radio) stacked on or near the product, but if it is required for your location to stack or use equipment that is adjacent to the product, all
must be verified to work and verification shall occur to ensure the product operates properly before conducting any procedures.
WARNING
No modification of this equipment is allowed.