Kimo KIGAZ 210 Software Operation Manual Download Page 1

Summary of Contents for KIGAZ 210

Page 1: ......

Page 2: ......

Page 3: ...Activate the group function 18 5 1 8 Select the alarm sound 19 5 1 9 Activate or deactivate Bluetooth function optional 19 5 2 Set the combustible 19 5 2 1 Modify a combustible 19 5 2 2 Add a combusti...

Page 4: ...nts 32 15 1 Perform a measurement of ambient temperature 32 15 2 Perform a measurement of flue gas temperature 32 15 3 Perform an ambient temperature measurement with the internal sensor 32 15 4 Perfo...

Page 5: ...ase 50 5 4 Edit an analysis report 51 6 Set the analyser 52 6 1 Change the combustibles 52 6 2 Change the references 52 6 3 Manage the operators 52 6 4 Modify the date and time 53 6 5 Sensors state 53...

Page 6: ...6...

Page 7: ...Go of memory The gas analyser is supplied in a transport case in ABS or metallic optional with a magnetic protective cover a differential pressure kit an interchangeable flue gas probe of 300 mm lengt...

Page 8: ...nal probes Pt100 temperature CH4 Flue gas probe connection P pressure tap P pressure tap Top view Bottom view Right side view Measurements menu Setting menu Information menu Function keys OK button ES...

Page 9: ...ture sensors Flue gas temperature is measured with a thermocouple integrated in the contact tip The connection to the analyser is made thanks to a single male connector compensated in temperature K th...

Page 10: ...the ticket This measurement must be performed with an opacity pump which is available as an accessory ref PMO Gas flow It is possible to measure the gas flow of an installation to compare it with a th...

Page 11: ...ferred to LHV Lower Heating Value and can exceed 100 The LHV is calculated by an algorithm developed by Kimo and its value must be considered as indicative CO O2 Concentration of CO in flue gases in m...

Page 12: ...ient temperature Pt100 1 3 DIN external probe From 50 to 250 C 0 1 C 0 3 of measured value 0 25 C 30 s Dew point temperature Calculated From 0 to 99 Ctd 0 1 C DHW temperature TcK external probe From 2...

Page 13: ...especting these safety instructions can lead to a loss of warranty The analyser should not to be used in and under water The analyser should not to be placed near or directly at open fire or heat The...

Page 14: ...e screen is displayed with measurement menu highlighted Insert the probe in the flue as show on the figure for at least 3 minutes Press OK The screen with the different possible measurements is displa...

Page 15: ...nd down arrows then press OK or create a customer see page 23 then select it The analyser displays the list of available boilers Select the boiler with up and down arrows then press OK or create a boi...

Page 16: ...nt with Max CO During the quick measurement it is possible to print a ticket at each step by pressing the Print function button or print the whole measurements when the analyser proposes it at the end...

Page 17: ...f auto off or OFF to deactivate it with the up and down arrows Press OK to validate 5 1 3 Change the language It is possible to select the language of the instrument French or English in standard A th...

Page 18: ...validate the line The keyboard disappears Press Esc to back to Setting instrument screen and validate modifications To go from lower case keyboard to upper case keyboard then to numeric keyboard press...

Page 19: ...ualified technician is authorized to modify them It is still possible to reset the factory parameters pressing the Factory parameters function key or with the LIGAZ 2 software see page 55 Setting scre...

Page 20: ...lowing parameters O2 CO Amb CO CH4 CO2 Qs losses excess air t lower efficiency s higher efficiency Tf flue gas temperature Ta ambient temperature Td dew point temperature Pr pressure P differential pr...

Page 21: ...chnician Setting menu is displayed Go to References with the keypad then press OK Select the parameter to modify with the keypad then press OK The keyboard appears Enter the parameter value with the k...

Page 22: ...ith the keypad then press OK 5 7 1 Units and values of measurement settings of the factory setting Temperature C Draft Pa Differential pressure mbar CO ppm CO2 NO ppm NOx ppm CH4 ppm CO O2 mg m3 NO O2...

Page 23: ...for both following lines of the address if necessary Go to Phone line with the keypad then press OK Enter customer s telephone number with the keypad and press OK to validate the numbers Press the Sa...

Page 24: ...f combustible for the boiler then press OK Press OK to tick the Air pulser burner box if the boiler has one Press the Save function key to save the boiler features The analyser backs to the list of bo...

Page 25: ...customer s inspections is displayed Press the Delete function key to delete an inspection Go to YES then press OK to delete the inspection or go to NO then press OK to cancel 8 4 Create a group Groups...

Page 26: ...OK To cancel go to NO and press OK or press Esc key 9 Perform measurement of a customised procedure Customised procedures created by the LOGAZ 2 software optional software when they are transferred in...

Page 27: ...AS H 19 21 mbar GAS B 25 mbar Butane Propane NF Pu KW x 2 166 Pu KW x 2 519 Pu KW x 0 651 Pu KW x 0 843 CE Pu KW x 1 927 Pu KW x 2 241 Pu KW x 0 579 Pu KW x 0 750 10 2 Perform a measurement of real fl...

Page 28: ...ork to 50 mbar then press OK A message indicating Please wait 10 minutes and a progress bar are displayed At the end of the 10 minutes results are displayed start pressure end pressure possible loss o...

Page 29: ...e flue gas probe on the analyser Turn on the analyser Go to the Measurement menu with the keypad then press OK The analyser performs the autozero remaining time and a progress bar display on the scree...

Page 30: ...13 2 Link the measurement to a customer When the measurement has been performed it is possible to link it to a customer registered in the analyser or to a customer unregistered in the analyser that m...

Page 31: ...he screen of measurement start of Max CO 14 2 Link the measurement to a customer When the measurement has been performed it is possible to link it to a customer registered in the analyser or to a cust...

Page 32: ...screen displays the NTC temperature detected by the analyser 15 4 Perform a Flow Return temperature measurement Connect 2 thermocouples Velcro type ref SKV 150 or lamella contact ref SKCT probes on th...

Page 33: ...ent screen 15 7 Print the results It is possible to print the results on the ticket Press the Print function button Results are printed on the ticket See page 17 for the printing configuration Perform...

Page 34: ...Press OK Press OK on Index 1 line Increment the index from 0 to 9 with the keypad Press OK Repeat the procedure for 2 and 3 indexes The analyser indicates the opacity indexes average on Average line a...

Page 35: ...een performed it is possible to link it to a customer registered in the analyser or to a customer unregistered in the analyser that must be created Press the Save function button The analyser displays...

Page 36: ...ith the keypad then press OK The measurement begins the analyser displays the CH4 level 18 3 Link the measurement to a customer When the measurement has been performed it is possible to link it to a c...

Page 37: ...d it is possible to link it to a customer registered in the analyser or to a customer unregistered in the analyser that must be created Press the Save function button The analyser displays the list of...

Page 38: ...Kimo Instruments Options of the analyser Type of sensor its serial number and its installation date To get to this menu From the home screen Go to Information menu with the keypad then press OK The in...

Page 39: ...Long life O2 5 years Annual CO H2 2 to 3 years Annual NO 2 to 3 years Annual CH4 3 to 5 years Annual 21 2 Replace the sensors Follow this procedure to replace a used sensor The analyser must be turned...

Page 40: ...rs 21 3 Replace the paper roll of the printer Follow this procedure to replace the paper roll of the printer Raise the upper part of the printer Put the paper roll in the required location Pull up sli...

Page 41: ...yser must be turned off Remove the protective cover Turn the analyser Unscrew the battery cover with a cross head screwdriver Remove the battery cover Disconnect the battery Connect the new battery Do...

Page 42: ...red in the device in O2 oxygen value measured in VF the theoretical volume of flue gas in m3 given in the table of coefficient NCV the net calorific value in MJ given in the table of coefficient CO2ma...

Page 43: ......

Page 44: ...ecommended minimum configuration For the proper functioning of the software the following minimum configuration is recommended Operating system Windows XP VISTA 7 Communication port USB 2 0 DVD player...

Page 45: ...nalyser and the instrument setting Setting allows to set different parameters autozero combustibles references Customers boilers deletion allows to delete customers and or boilers saved in the analyse...

Page 46: ...l bar The software recovers inspections contained in the analyser This operation can take a few minutes At the end of the recovery the following window is displayed Tick the boxes corresponding to the...

Page 47: ...are not accessible The customer list created on the device are displayed on the left of the screen To add a customer click on the button The fields become accessible Fill in the different fields The N...

Page 48: ...ol bar button or on Database menu then choose Boilers The following window opens All the fields type model power are inaccessible and the Browse button is inactive To add a boiler click on the button...

Page 49: ...ist then click on the button If it is not accessible check the device connection A message is displayed to confirm the transfer Click on OK To download boilers from the analyser to the software databa...

Page 50: ...hen on Database settings The database configuration window opens Click on the button and define the save folder path Select the recurrence that is the automatic registration of the database Select the...

Page 51: ...n the menu on the left of the screen click on the required boiler for the report edition Click on the button The analysis report is displayed Use the button to zoom in or zoom out on the report Click...

Page 52: ...combustibles This window allows to change the list of combustibles programmed in the analyser It also allows to modify features of each combustibles KIGAZ settings window is open on Combustibles or c...

Page 53: ...nt KIGAZ settings window is open on Information or click on the button The following information is displayed Product code ex KIGAZ 310 Serial number ex 00 01 2057 Firmware version ex 0 92 Build ex 12...

Page 54: ...ver 7 Copy the address and the port in the Address line of the connection window as follows 00 00 000 00 0000 Enter the user name and the corresponding password data entered at the starting of the com...

Page 55: ...following window is displayed Fill in the different fields Email address of the outgoing mail e mail address of the sender p durand orange fr Outgoing mail server server address used to send e mail e...

Page 56: ...ustomer to delete then click on the button between Customer and Boiler columns A message of confirmation is displayed Click on Yes to confirm the deletion To delete a boiler on the analyser Click on t...

Page 57: ...atically customers and boilers in the following cases If the same customer is present in the analyser and the software Customers database but different boilers are associated to this customer the soft...

Page 58: ...he PC and KIGAZ box Click on the OK button To update the information Tick the Synchronise data of the PC and KIGAZ box Tick the boxes of data to keep Click on OK button To do the same actions on the f...

Page 59: ...Once returned to KIMO required waste collection will be assured in the respect of the environment in accordance to guidelines relating to WEEE...

Page 60: ...nique avec les ordures m nag res Renvoyez le chez KIMO au terme de sa dur e d utilisation Conform ment aux directives europ ennes relatives aux DEEE nous assurons une collecte distincte pour un traite...