4
1.2 Protection in the transportation
Before the initial use, please refer to the following drawing to
take off the protection screw. When packing the scale for
transportation, install the protection screw on the scale in order
to avoid any damage to sensitive components.
1.3 Leveling the Scale
The scale is equipped with a level indicator on the left bottom
side of the front panel and four adjustable leveling feet. Adjust
the leveling feet until the bubble appears in the center circle of
the indicator. Then lock the feet in place using the lock nuts.
Repeat this procedure each time the scale is moved to a new
location.
1.4 Turn on Scale
Do not turn on scale with anything on the platform. The
switch is located on the right side of the scale. The scale will
start to count down from nine to zero and will display the max
capacity of the scale in the display. The scale is then ready
for use.
Allow the scale15-30 minutes to warm up prior to use.