38
TOS7200
Initialize
To initialize, turn ON the POWER switch with the SHIFT key held down. This clears the contents
of memory and restores tester settings to the factory default settings. (Press the SHIFT key
until the timer’s 7-segment LED starts to blink.)
The panel settings restored following initialization are given in Table below. For settings for test
conditions aside from the test-voltage set value, the same value is stored in panel memories 0 to 9.
Settings for test conditions
Test voltage settings
System setting
Initializing clears the contents of the panel settings stored in memory. Before initializing,
make sure no data still required is contained in memory.
Item
Set value
Test voltage (VOLTAGE)
See
in the
following table.
Lower resistance (LOWER)
1.00 MΩ
Upper judgment (UPPER ON)
ON
Upper resistance (UPPER)
100 MΩ
Timer (TIMER ON)
ON
Timer set value (TIMER)
0.5 s
Wait time (WAIT TIME)
0.3 s
Panel memory number
Test voltage set value
0
10 V
1
25 V
2
50 V
3
100 V
4
125 V
5
250 V
6
500 V
7 to 9
1000 V
Item
Set value
Double action (DOUBLE ACTION)
OFF
Momentary (MOMENTARY)
OFF
Fail mode (FAIL MODE)
OFF
Pass hold (PASS HOLD)
OFF
Average count (AVG)
1 (fixed)
Auto-range (AUTO RANGE)
ON
Lower judgment function (LOWER ON)
ON
Communications rate (SPEED)
19200 bps
Buzzer volume (BUZZER)
5