TOS5200
51
5
Withst
andin
g
V
o
lt
age
T
e
st
Finishing a Test
To stop a test
Press STOP.
When a test finishes
A test will stop under one of the following conditions.
a. If the test time elapses (when TIMER is set to ON).
b. If a current greater than or equal to the upper limit (U-FAIL) or less than or equal to the
lower limit (L-FAIL) is measured.
c. If you press STOP.
After a test finishes, the DANGER LED turns off, and the high voltage output is turned off.
If a test finishes under condition a or b given above, the judgment result is displayed on the
screen.
If the test time (Timer) has been set
p. 41 , p. 42
After the test time elapses, the PASS LED lights in green, and the test finishes. The PASS
LED remains lit in green for the length of time specified by Pass Hold. The READY LED then
lights in light white, and the TOS5200 switches to the READY state.
For details on how measured values are judged, see “About Judgment”.
The operations that the TOS5200 performs after testing finishes for each judgment condition
are shown below.
STOP
A
kV
M
ȍ
mA
CONFIG
RISE
TEST
RMS
AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
s
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
Measured voltage
Measured current Test time
Example: Display of a PASS judgment
after testing finished
See
A
kV
M
ȍ
mA
CONFIG
RISE
TEST
RMS
AVE
MORE CAL RMT
MT
LIMIT
50Hz
60Hz
UPPER
LOWER W COMP
1
2
3
READY
TEST
PASS
FAIL
PROTECTION
s
See
Operation
PASS
U-FAIL
L-FAIL
Display
PASS LED lights in green.
Measured value remains
displayed for the length of
time specified by Pass
Hold.
FAIL LED (yellow) and
“UPPER” lights.
FAIL LED (yellow) and
“LOWER” lights.
Buzzer
Sounds for 50 ms.
1
1
The length of time that the buzzer sounds when a PASS judgment occurs is fixed to 50 ms. The
specified Pass Hold time has no effect.
Sounds until STOP is pressed.
SIGNAL I/O
connector
The PASS signal is gener-
ated for the length of time
specified by the Pass Hold
setting.
The U-FAIL signal is gener-
ated until the FAIL judgment
is cleared.
The L-FAIL signal is gener-
ated until the FAIL judgment
is cleared.
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......