If the test result is PASS
When the preset test time elapses, the tester will make a PASS judgment. If the test
result is a PASS, the following actions will take place.
♦
The tester terminates the test and cuts off the test voltage.
♦
The “TEST” message disappears.
♦
The TEST signal of the SIGNAL I/O connector is turned OFF.
♦
The DANGER lamp turns OFF. (It does not go out as long as a residual test
voltage remains in the output circuit.)
♦
The H.V ON signal of the SIGNAL I/O connector is turned OFF. (It is not
turned off as long as a residual test voltage remains in the output circuit.)
♦
The “PASS” message appears.
♦
The buzzer sounds.
♦
The PASS signal of the SIGNAL I/O connector is turned ON.
The Vacuum Fluorescent Display in this case will be as follows:
When in the normal state, the tester will automatically reset to the READY status
(status that the “READY” message appears) within approximately 0.2 seconds from
the PASS status and the following actions will take place.
♦
The “PASS” message disappears.
♦
The buzzer stops.
♦
The PASS signal of the SIGNAL I/O connector is turned OFF.
• The tester will not automatically reset to the READY status when in any of the
following cases:
♦
When the PASS HOLD switch of the TEST MODE switches on the rear panel has
been set to ON. In this case the PASS status is held continuously. To reset it, press
the STOP switch. (For details, refer to section 6.5, “Settings for Special Test
Modes” on Page 6-29 .)
♦
As long as the cause of the PROTECTION status remains. (For details, refer to
[32] PROTECTION message in section 4.2, “Display Items” on Page 4-18 .)
♦
When the STOP switch is pressed (including the STOP switch of the remote con-
trol box).
♦
When the START switch remains in the pressed state (including the START
switch of the remote control box).
PROTECTION
READY
TEST
PASS
AC
kV
mA
FAIL
DC
s
CURRENT
TIME
UPPER LOWER
KEYLOCK
REMOTE
TIMER ON
LOWER ON
UPPER
LOWER
RANGE
kV
VOLTAGE
Lights up
NOTE