74
TOS3200
Program Tests
A sequence program allows various tests to be executed consecutively by changing the test
conditions of the PCC or TC and the test lead connections.
For example, the following five tests can be executed consecutively by simply changing the
test lead connections between each test (interval), and the result can be stored.
Up to 100 sequence programs can be created. You can assign a name to each program
using a number between 00 and 99 and up to 12 characters.
Each sequence program can consist up to 100 single tests (100 steps). However, the total
number of steps of all programs is 500. For example, up to five sequence programs consist-
ing of 100 steps each can be created.
Press the AUTO key to display the program test screen 1/2 (AUTO1/2).
00 PCC measurement
01 TC measurement between the enclosure and earth
02 TC measurement between two enclosures
03 TC measurement between the enclosure and power line (live)
04 TC measurement between the enclosure and power line (neutral)
AUTO 1 / 2
00–TC•E
NC
P
E
––P
NRM
•NORM––– 1s
01–TC•E
NC
E
NC
–P
NRM
•NORM––– 1s
–––END
PRG
01:--UNTITLED--
TOTAL: 2
READY
AUTO 1 / 2
–––END
PRG
00:--UNTITLED--
TOTAL: 0
Program name (up to 12 characters)
Program number (00 to 99)
Screen title
The screen when a program in which no
steps are configured is selected
Program test screen
consisting of two steps
Summary of the test conditions
Total number of steps of the selected
program
Program test ready status
Displayed only if the selected
program number contains steps.
Step number
(00 to 99)
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......