44
PMC-A
・
After inserting the wire, check that it does not come
loose.
・
Be s
u
re the exposed section of the wire does not
to
u
ch the chassis or other wires s
u
ch as the wire of
the adjacent terminal.
Connection and setup procedure
1.
T
u
rn off the PO
W
ER switch and OUTPUT switch.
2.
Connect the cables as shown in Fig. 4-3
For the handling of the SENSING terminal, see section Fig. 4-2
3.
T
u
rn on the REMOTE SE
N
SI
N
G switch.
Fig. 4-3 Connection of the sensing cable
・
For sensing cables, use cables with a higher
voltage rating than the isolation voltage of the
power supply.
・
B
u
rno
u
t may occ
u
r in the load. If the sensing wires
come loose, the o
u
tp
u
t voltage across the load can-
not be stabilized and may ca
u
se excessive voltage
to be applied to the load. Sec
u
rely connect the
sensing wires s
u
ch as by
u
sing crimp terminals.
CAUTION
REMOTE SENSING
OFF ON -S +S
+
+
Load
+S
Load ca
b
les
To OUTPUT
terminals
T
w
ist
-S
WARNING
CAUTION