The PL02-TOS warning light unit indicates that a Kikusui withstanding
voltage tester, withstanding voltage and insulation resistance tester, or
electrical safety analyzer (hereafter referred to as the Tester) that has a
STATUS OUT (24 V) terminal is in a test state.
For your safety, be sure to check the warning light illumi-
nates before using it. Do not leave or use the PL02-TOS
with the cover removed or when the cover is broken.
Names of Parts
■
Connecting to the connector
1
Remove 10 mm of coating from
the tip of the wires.
2
Insert a flat-blade screwdriver
into A, and open B.
3
Insert the wire into B. Take care
to prevent the coating from becoming caught in B.
4
Gently pull on the wire to confirm that it is securely connected.
OPERATION MANUAL
Warning Light Unit
PL02-TOS
Part No. Z1-002-690, IA002883
Dec 2018
WARNING
Cover
Illuminates or blinks in red when power is input.
BLINK ON/OFF
Switch used to select illuminating or blinking of the warning
light. Turn the switch ON to set the warning light to blink.
AC/DC24V
Connector for the input power. There is no polarity.
A
B
Shank diameter: ø3; tip width: 2.6 mm
10 mm
Applicable wires
• Wire length
Less than 3 m
• Solid wire
ø0.4 to ø1.2 (AWG26 to 16)
• Stranded wire
0.3 mm
2
to 1.25 mm
2
(AWG22 to 16)
Element wire diameter ø0.18 or more