This section introduces troubleshooting measures. Typical symptoms are listed. Check
whether any of the items listed below apply to your case. In some cases, the problem can be
solved quite easily.
If none of the items apply to your case, we recommend that you initialize the PBZ to its
factory default settings. If the remedy does not correct the problem, Contact your Kikusui
agent or distributor.
See
The power does not turn on.
Symptom
Items to check and possible remedy
The PBZ does not operate
when the POWER switch is
turned on.
Is the power cord broken?
Replace the power cord with the same type of power cord.
An error number (E901) is
displayed after the self-test.
An error was detected in the internal system.
Restart the PBZ.
See
Unable to carry out panel operations.
Symptom
Items to check and possible remedy
Unable to operate the panel
keys.
Is the key lock icon displayed on the screen?
Release the panel operation lock.
Is the RMT icon displayed in the screen’s status display area?
Is the PBZ being controlled via the RS232C, GPIB, USB, or LAN interface?
To control the PBZ from the panel, press the LOCAL key to switch to local mode.
The PBZ does not switch to
local mode even when I press
the LOCAL key.
Was a local lockout (LLO) command sent through the communication interface?
Use a communication command to clear the local lockout (LLO) command. For
details, see the communication interface manual.
–
See
Output does not turn on.
Symptom
Items to check and possible remedy
Output does not turn on when
the POWER switch is turned
on.
Did you press the OUTPUT key?
Press the OUTPUT key.
Have you set the PBZ to turn output on when the POWER switch is turned on?
In the CONFIG settings, set POWER ON>OUTPUT to ON.
Output does not turn on even
when I press the OUTPUT
key.
Are you using an external contact to turn output on and off?
Use the external contact to turn output on.
Is the external control logic setting for turning output on and off correct?
In the CONFIG settings, set EXT OUTPUT ON/OFF>POLARITY to HIGH (output
turns on when the contact enters the break state).