Troubleshooting
This section introduces troubleshooting measures. Typical symptoms are listed. Check whether any of the
items listed below apply to your case. In some cases, the problem can be solved quite easily.
If none of the items apply to your case, we recommend that you initialize the product to its factory default
settings
. If following the remedy does not solve your problem, contact your Kikusui agent or distrib-
utor.
Condition
Possible cause
Remedy
See
Nothing appears on the display
when the POWER switch is
turned on.
Rated voltage is not applied to
AC INPUT.
Check that the power cord is not
broken and that the connection at
the AC INPUT inlet is secure.
“SYSTEM ERROR” is shown in
the display.
A system error is occurring.
Turn the POWER switch off and
then turn it back on.
–
Keys do not work.
Key lock is activated.
Press KEY LOCK (SHIFT +
ENTER) to release the key lock.
–
The product is under remote
control. (“REMOTE” is shown
in the display.)
Press LOCAL (SHIFT+ESCAPE) to
switch to local mode allowing you to
control the product from the front
panel.
–
Measurement is not possible.
The trigger source is set to
EXTERNAL.
If you do not want to apply trigger
signals manually, set the trigger
source to INTERNAL.