9. The dispenser is ready for operation when a red ray of light will be displayed in the dosing space
between the dosing nozzle and the waste fluid drain. The dispenser is now ready for use.
10. After the installation is complete, check that the dispenser is dispensing hand disinfectant properly
(ref. 3.2.3).
NOTE! Recycle or dispose of the used hand disinfectant canister in an appropriate manner. For more
details on safe disposal, contact your local KiiltoClean representative or the respective environmental
authority.
5.3 Troubleshooting guide
Description
Repair procedure
The indicator lights are all out.
Check that the power transformer is properly connected to the
dispenser and to the electrical outlet and that the electrical outlet is
functioning.
If the dispenser fails to operate after these steps, please contact
customer service. The contact information is listed at the end of this
user manual.
The dispensing pump is operating, but no
hand disinfectant is dispensed.
Check that the suction line is not bent or clogged preventing the
hand disinfectant from flowing.
If the dispenser fails to operate after these steps, please contact
customer service. The contact information is listed at the end of this
user manual.
The dispenser does not react when hands are
placed in the dispensing opening.
Check that the power transformer is properly connected to the
dispenser and to the electrical outlet and that the electrical outlet is
functioning.
If the dispenser fails to operate after these steps, please contact
customer service. The contact information is listed at the end of this
user manual.
The hand disinfectant splashes outside the
dispenser.
Instruct people to keep their hands in the dispensing opening
during the dispensing.
Check the amount of dispensed hand disinfectant. The correct
amount has been programmed at the time of the delivery.
If the amount dispensed does not match with the delivery
specifications, please contact customer service. The contact
information is listed at the end of this user manual.
The waste fluid canister does not collect the
hand disinfectant.
Empty the waste fluid canister as per the instructions in this
user manual