1
Calibration Procedures
46
Keysight U3606B Service Guide
4-wire resistance gain verification test
1
Connect the calibrator to the front panel (red) and
LO
(black) input
terminals and
(red) and (black) output terminals (see
setup for 4-wire resistance gain verification,”
2
Press
[
]
to select the resistance function.
3
Select each range in the order shown below. Provide the indicated input
resistance. Compare the measurement results to the appropriate test limits in
(be certain to allow for appropriate source settling when using the
Fluke 5520A).
Table 1-7
2-wire resistance gain verification test
Input resistance
Range
Error from nominal 1-year
100
Ω
100
Ω
±0.058
Ω
[a]
[a] Specifications stated are for 2-wire resistance measurements using the Null math operation.
Without null, add an additional error of 0.2
Ω
.
1 k
Ω
1 k
Ω
±0.00055 k
Ω
10 k
Ω
10 k
Ω
±0.0055 k
100 k
Ω
100 k
Ω
±0.055 k
Ω
1 M
Ω
1 M
Ω
±0.00065 M
Ω
10 M
Ω
10 M
Ω
±0.0255 M
Ω
100 M
Ω
100 M
Ω
±2.005 M
Ω
NOTE
For the
1
Ω
and 10
Ω
range, please perform calibrator zeroing before testing the
resistance values of the device-under-test (DUT) to ensure that the 5520A is
maintained with a zero cal. every 12 hours ± 1
o
C.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...