Calibration Procedures
1
Keysight U3606B Service Guide
45
AC current verification test
1
Connect the calibrator to the front panel
I
(red) and
LO
(black) input terminals
(see
Figure 1-6, “Test setup for DC current and AC current gain verification,”
2
Press
[
]
to select the AC current function. The AC annunciator will turn on
according to your selection.
3
Select each range in the order shown below. Provide the indicated input
current and frequency. Compare the measurement results to the appropriate
test limits in
. (Be certain to allow for appropriate source settling
when using the Fluke 5520A.)
2-wire resistance gain verification test
1
Connect the calibrator to the front panel
(red) and
LO
(black) input
terminals (see
Figure 1-4, “Test setup for 2-wire resistance gain verification,”
2
Press
[
]
to select the resistance function.
3
Select each range in the order shown below. Provide the indicated input
resistance. Compare the measurement results to the appropriate test limits in
(be certain to allow for appropriate source settling when using the
Fluke 5520A).
Table 1-6
AC current gain verification test
Input current
Input frequency
Range
Error from nominal 1-year
1 mA
1 kHz
10 mA
±0.015 mA
10 mA
1 kHz
10 mA
±0.06 mA
10 mA
1 kHz
100 mA
±0.15 mA
100 mA
1 kHz
100 mA
±0.6 mA
0.1 A
1 kHz
1 A
±0.0015 A
1 A
1 kHz
1 A
±0.006 A
1 A
1 kHz
3 A
±0.008 A
3 A
1 kHz
3 A
±0.018 A
I
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...