![Keysight U3606A User'S And Service Manual Download Page 230](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516230.webp)
6
Calibration Procedures
Adjustments procedures
208
U3606A User’s and Service Guide
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input (see
“Entering adjustment values” on
7
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the capacitance gain adjustments using the
gain verification test” on page 172
N O T E
If the zero offset adjustment procedure has been recently performed prior
to the capacitance gain calibration procedure, the adjustment item
“OPEn” can be omitted.
N O T E
Always complete tests in the same order as shown in
.
Table 6-9
Capacitance gain adjustment
Adjustment item
Input capacitance
OPEn
Input terminals open (remove any test leads or
shorting plugs from the input terminals)
0.400 nF
0.4 nF
1.000 nF
1 nF
O U T
S B Y
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...