Keysight U3020A User’s and Service Guide
13
General Performance
The Test Set performance is specified for a standalone instrument.
Power requirements
– 100/120/220/240 VAC (50/60 Hz)
– The instruments can operate with mains supply voltage fluctuations up to ± 10% of the nominal
voltage.
– Air conditioning equipment (or other motor–operated equipment) should not be placed on the
same ac line that powers the Test Set.
– U3022AS26 maximum power is 50 W.
Environmental tests
The U3020AS26 complies with all applicable safety and regulatory requirements for the intended
location of use.
– Operating environment (indoor use)
– Operating temperature 0 to 55 °C
– Storage temperature: -40 to +70 °C
– Operating altitude: 0 to 2000 meters (~ 6,562 feet)
– The instrument can safely operate in a relative humidity of 80% for temperatures to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C.
Dimensions and space requirements
Standard installation of the U3020AS26 includes configuration and installation on a customer
provided lab bench, or table top of adequate size and strength.
CAUTION
This product has an autoranging line voltage input. Be sure the supply voltage is within the
specified range.
Table 1
Test set dimensions
Item
Length/Weight
Width
42.5 cm (16.73 in)
Depth
58 cm (22.83 in)
Weight
9 kg (19.84 lb)
Summary of Contents for U3020AS26
Page 1: ...Keysight U3020AS26 Switching Test Set User s and Service Guide ...
Page 4: ...4 Keysight U3020A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 8: ...8 Keysight U3020A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U3020A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 40: ...40 Keysight U3020A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...