Table 7-5
Calibration tests
Step
Test function
Cal range
Input
U1401B
1.
Turn the rotary switch to
mV
. Press
to
select DC.
50 mV
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
45 mV
45 mV
45.000 mV
500 mV
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
450 mV
450 mV
450.00 mV
Turn the rotary switch to
. Press
to
select DC.
5 V
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
4.5 V
4.5 V
4.5000 V
50 V
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
45 V
45 V
45.000 V
500 V
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
450 V
450 V
450.00 V
1000 V
Short
Dual banana plug with
copper wire shorted
between 2 terminals
Short
900 V
900 V
900.0 V
2.
Turn the rotary switch to
mV
. Press
to
select AC.
50 mV @ 1 kHz
5 mV
5 mV
5.000 mV
45 mV
45 mV
45.000 mV
500 mV @ 1 kHz
50 mV
50 mV
50.00 mV
450 mV
450 mV
450.00 mV