background image

Calibration Procedures

1

Keysight U1281A/U1282A Service Guide

51

Vsense (non-contact voltage) calibration 

1

Turn rotary switch to 

 position with the 

shown on the display. 

2

Connect a dual open type banana plug into the output voltage terminals of the 

calibrator.

3

Ensure the 

EARTH

 button of calibrator is on.

4

Hold the multimeter so that the top of the multimeter is vertically and 

horizontally centered and in contact with the banana plug’s Hi terminal.

5

Set the source output to 10 V/55 Hz.

6

Press 

 to start the calibration. If the calibration is successful, 

 will 

be displayed.

7

Set the source output to 30 V/55 Hz.

8

Press 

 to start the calibration process. 

9

Set the calibrator to standby mode when the calibration is done.

See 

Figure 1-8

 for the multimeter display.

NOTE

– Ensure the multimeter is REMOVED from the holster prior to performing this 

calibration.

– Keep the multimeter away from fluorescent lights, dimmable lights, motors, 

and other electrical noise sources when running the tests. These electrical 

noise sources can invalidate the calibration.

– It may be necessary in 

step 4

 and 

step 8

 below to slightly adjust the position 

of the multimeter for maximum signal strength.

V

Hz

Save

Hz

Save

Summary of Contents for U1281A

Page 1: ...Keysight U1281A U1282A Handheld Digital Multimeter Service Guide ...

Page 2: ...hts customarily provided to the public to use modify reproduce release per form display or disclose commercial computer software or commercial com puter software documentation No additional government requirements beyond those set forth in the EULA shall apply except to the extent that those terms rights or licenses are explicitly required from all providers of commercial computer software pursu a...

Page 3: ...of the instrument Direct current DC Alternating current AC Caution risk of electric shock Caution risk of danger refer to this manual for specific Warning or Caution information Earth ground terminal Equipment protected throughout by double insulation or reinforced insulation Category III 1000 V overvoltage protection Category IV 600 V overvoltage protection CAT III 1000 V CAT IV 600 V ...

Page 4: ... particular attention to the insulation surrounding the connectors Inspect the test leads for damaged insulation or exposed metal Check the test leads for continuity Replace damaged test leads before you use the multimeter Do not operate the multimeter around explosive gas vapor or wet environments Do not apply more than the rated voltage as marked on the multimeter between terminals or between te...

Page 5: ...ttery as soon as the low battery indicator appears and flashes Only probe assemblies with Measurement Category III or IV ratings should be used for mains measurements CAUTION Disconnect circuit power and discharge all high voltage capacitors before testing resistance continuity diodes or capacitance Use the proper terminals function and range for your measurements This multimeter is for use at alt...

Page 6: ...les are measurements on household appliances portable tools and similar equipment Measurement CAT III Measurements performed in the building installation Examples are measurements on distribution boards circuit breakers wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and equipment for industrial use and some other equipment including stationary mot...

Page 7: ...al environmental requirements for this instrument Environmental condition Requirement Temperature Operating condition 20 C to 55 C 0 to 80 RH non condensing Storage condition 40 C to 70 C 0 to 80 RH without batteries Humidity Full accuracy up to 80 RH relative humidity for temperature up to 30 C decreasing linearly to 50 RH at 55 C Altitude Up to 3000 meters Pollution degree 2 ...

Page 8: ...tic Compatibility EMC compliances Safety compliance IEC EN 61010 1 2010 3rd Edition IEC EN 61010 2 033 2012 First Edition CAN CSA C22 2 No 61010 1 12 CAN CSA C22 2 No 61010 033 12 ANSI UL Std No 61010 1 12 ANSI UL Std No 61010 033 12 EMC compliance IEC 61326 1 2012 EN 61326 1 2013 ICES NMB 001 Issue 4 June 2006 AS NZS CISPR 11 2004 ...

Page 9: ...th the WEEE Directive 2002 96 EC marking requirement This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste The RCM mark is a registered trademark of the Spectrum Management Agency of Australia This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992 This s...

Page 10: ...nd Control Instrument product The affixed product label is as shown below Do not dispose in domestic household waste To return this unwanted instrument contact your nearest Keysight Service Center or visit http about keysight com en companyinfo environment takeback shtml for more information Sales and Technical Support To contact Keysight for sales and technical support refer to the support links ...

Page 11: ...ct category 10 Sales and Technical Support 10 1 Calibration Procedures Calibration Overview 18 Closed case calibration 18 Keysight calibration services 18 Calibration interval 18 Recommended adjustment 19 Recommended Test Equipment 20 Basic Operating Tests 21 Backlight test 21 Display test 21 Input warning test A terminal 22 Input warning test μ mA terminal 24 Calibration Process 26 Test Considera...

Page 12: ...tions 45 Valid adjustment input values 46 Vsense non contact voltage calibration 51 Adjustment procedure 53 Exiting the adjustment mode 54 Calibration Count 55 Calibration Error Codes 57 2 Service and Maintenance Troubleshooting 60 Fuse Replacement 61 Returning the Multimeter for Service 64 Replaceable Parts 65 To order replaceable parts 65 Types of Service Available 66 Extended service contracts ...

Page 13: ...ibration security code entry display 38 Figure 1 5 Multimeter display when unsecuring the multimeter 40 Figure 1 6 Multimeter display when changing the calibration security code 42 Figure 1 7 Multimeter display when resetting the calibration security code to its factory default 44 Figure 1 8 Multimeter display for Vsense non contact voltage calibration 52 Figure 1 9 Multimeter display for calibrat...

Page 14: ...14 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 15: ...9 AC current A performance verification tests 32 Table 1 10 AC current μA mA performance verification tests 33 Table 1 11 Diode performance verification test 33 Table 1 12 Capacitance performance verification tests 34 Table 1 13 Resistance performance verification tests 34 Table 1 14 Conductance performance verification test 35 Table 1 15 Frequency sensitivity performance verification test 35 Tabl...

Page 16: ...16 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 17: ...perating Tests 21 Calibration Process 26 Test Considerations 26 Performance Verification Tests 27 Calibration Security 37 Unsecuring the Multimeter for Calibration 38 Using the Front Panel for Adjustments 45 Calibration Count 55 Calibration Error Codes 57 This chapter helps you to verify the multimeter performance and to make adjustments where necessary ...

Page 18: ...e stored in the nonvolatile memory until the next calibration adjustment is performed The nonvolatile EEPROM calibration memory is retained even when the power is switched off Keysight calibration services When your multimeter is due for calibration contact your local Keysight Service Center to enquire about recalibration services Calibration interval A 1 year interval is adequate for most applica...

Page 19: ... performance This will ensure that the multimeter will remain within the specifications for the next calibration interval This criterion for the re adjustment provides the best long term stability Performance data are measured during the Performance Verification Tests on page 27 but the multimeter is not guaranteed to remain within these limits unless the adjustments are performed See Calibration ...

Page 20: ... AC voltage Fluke 5520A 5522A 20 of the multimeter s 1 year specifications DC current Fluke 5520A 5522A 20 of the multimeter s 1 year specifications AC current Fluke 5520A 5522A 20 of the multimeter s 1 year specifications Capacitance Fluke 5520A 5522A 20 of the multimeter s 1 year specifications Conductance Fluke 5520A 5522A 20 of the multimeter s 1 year specifications Frequency Fluke 5520A 5522A...

Page 21: ...is required if the multimeter fails the basic operating tests Backlight test Press to test the backlight It will momentarily turn the backlight on and off Display test Press and turn on the multimeter by turning the rotary switch from to any other position to view all annunciators of the display Compare the display with the example in Figure 1 1 Figure 1 1 Display screen Hold ...

Page 22: ...t to the corresponding location The display indicates a flashing annunciator until the test lead is removed from the input terminal or until the rotary switch is set to the function Perform the following steps to replicate the input terminal warning test 1 Turn the rotary switch to the position 2 Insert the red test lead to the input terminal 3 Insert the black test lead to the input terminal A NO...

Page 23: ...1281A U1282A Service Guide 23 Figure 1 2 Example of A Err for wrong terminal input A mA COM V V A V mV mV U1282A True RMS Multimeter mA Vsense Peak Vsense Dual Auto FUSED 440mA CAT III 1000V 600V CAT IV MAX 10A MAX FUSED OUT OUT MHz RMT ...

Page 24: ...igure 1 3 The alert beep tone will continue to beep until the test lead is removed from the terminal or until the rotary switch is set to the function Perform the following steps to replicate the terminal input warning test 1 Turn the rotary switch to the position 2 Insert the red test lead to the input terminal 3 Insert the black test lead to the input terminal mA NOTE Before conducting this test...

Page 25: ...1281A U1282A Service Guide 25 Figure 1 3 Example of μAErr for wrong terminal input A mA COM V V A V mV mV U1282A True RMS Multimeter mA Vsense Peak Vsense Dual Auto FUSED 440mA CAT III 1000V 600V CAT IV MAX 10A MAX FUSED OUT OUT MHz RMT ...

Page 26: ...urned off 7 Record the new security code and calibration Test Considerations For optimum performance all procedures should comply with the following recommendations The performance verification test or adjustment should be performed under laboratory conditions where the ambient temperature can be controlled The multimeter should be put under the laboratory environment for at least 1 hour Ensure th...

Page 27: ... the multimeter The acceptance test results should be compared against the 1 year test limits After acceptance you should repeat the performance verification tests at every calibration interval If the multimeter fails the performance verification tests adjustment or repair is required NOTE Ensure that you have read the Test Considerations on page 26 before running the performance verification test...

Page 28: ...on Range 5520A 5522A output Error from nominal 1 year U1281A U1282A DC mV mV Turn the rotary switch to the position 60 mV a 60 mV 0 040 mV 0 040 mV 600 mV a 600 mV 0 20 mV 0 20 mV a The accuracy is specified after the Null function is used to subtract the thermal effect by shorting the test leads before measuring the signal mV ...

Page 29: ...6 V 10 kHz 0 0445 V 0 0445 V 6 V 20 kHz 0 0960 V 0 0960 V 6 V 100 kHz N A 0 3160 V 60 V 30 V 20 Hz 0 360 V 0 360 V 60 V 45 Hz 0 205 V 0 205 V 60 V 1 kHz 0 205 V 0 205 V 60 V 10 kHz 0 445 V 0 445 V 60 V 20 kHz 0 960 V 0 960 V 60 V 100 kHz N A 3 160 V 600 V 30 V 20 Hz 0 90 V 0 90 V 600 V 45 Hz 2 05 V 2 05 V 600 V 1 kHz 2 05 V 2 05 V 600 V 10 kHz 4 45 V 4 45 V 300 V 20 kHz 5 10 V 5 10 V 320 V 30 kHz ...

Page 30: ...V 1 kHz 0 205 mV 0 205 mV 60 mV 10 kHz 0 445 mV 0 445 mV 60 mV 20 kHz 0 960 mV 0 960 mV 60 mV 100 kHz N A 3 160 mV 600 mV 600 mV 20 Hz 6 60 mV 6 60 mV 600 mV 45 Hz 2 05 mV 2 05 mV 600 mV 1 kHz 2 05 mV 2 05 mV 600 mV 10 kHz 4 45 mV 4 45 mV 600 mV 20 kHz 9 60 mV 9 60 mV 600 mV 100 kHz N A 31 60 mV mV Esc Shift Table 1 6 LPF performance verification test Test function Range 5520A 5522A output Error f...

Page 31: ...vely use the Null function to zero the thermal effect before proceeding with a low current measurement or checking the current offset mA CAUTION Connect the calibrator outputs to the multimeter s A and COM terminals before applying 6 A and 10 A Table 1 8 DC current A performance verification tests Test function Range 5520A 5522A output Error from nominal 1 year U1281A U1282A DC current A Turn the ...

Page 32: ...A 0 0465 A 10 A b 1 A 20 Hz a 0 050 A 0 050 A 10 A 45 Hz 0 085 A 0 085 A 10 A 1 kHz 0 085 A 0 085 A 2 9 A 5 kHz 0 074 A 0 074 A a Input current is 3A rms due to verification source b Current can be measured from 1 A to 10 A continuously with an additional buffer of 0 3 If the measurement of the signal is between 10 A to 20 A current can be measured for 30 seconds maximum After measuring 10 A curre...

Page 33: ... 385 mA 0 385 mA 600 mA a 440 mA 20 Hz 4 80 mA 4 80 mA 440 mA 45 Hz 2 89 mA 2 89 mA 440 mA 1 kHz 2 89 mA 2 89 mA a Current can be measured from 30 mA to 440 mA continuously and 440 mA to 600 mA for 20 hours maximum DC ranging from 60 mA to 440 mA have a thermal effect of 0 35 µA mA Cool down the multimeter for 6 seconds if 100 mA is applied and at least 3 minutes if 600 mA is applied Alternatively...

Page 34: ...F 10 mF 10 mF 0 105 mF 0 105 mF 100 mF 100 mF 2 60 mF 2 60 mF Table 1 13 Resistance performance verification tests Test function Range 5520A 5522A output Error from nominal 1 year U1281A U1282A Resistance Turn the rotary switch to the position 60 Ω a 60 Ω N A 0 110 Ω 600 Ω a 600 Ω 0 40 Ω 0 40 Ω 6 kΩ 6 kΩ 0 0032 kΩ 0 0032 kΩ 60 kΩ 60 kΩ 0 032 kΩ 0 032 kΩ 600 kΩ 600 kΩ 0 32 kΩ 0 32 kΩ 6 MΩ 6 MΩ 0 00...

Page 35: ...ll function is used on an open test lead Range Table 1 15 Frequency sensitivity performance verification test Test function Range 5520A 5522A output Error from nominal 1 year U1281A U1282A Frequency sensitivity While the rotary switch is in the press 99 999 Hz 90 Hz 0 25 V 0 021 Hz 0 021 Hz V Hz Save Table 1 16 Duty cycle performance verification test Test function Range 5520A 5522A output Error f...

Page 36: ... internal reference To perform the measurement connect one end of the K type thermocouple with miniature TC connector on both ends to the 5520A TC output and the other end to the multimeter via a TC to banana adapter Allow at least 1 hour for the multimeter to stabilize Table 1 18 Frequency counter performance verification test Test function Range 33522B output Error from nominal 1 year U1281A U12...

Page 37: ...t unsecure it by entering the correct security code See Unsecuring the Multimeter for Calibration on page 38 The security code is set to when the multimeter is shipped from the factory The security code is stored in nonvolatile memory and does not change when power has been turned off The security code may contain up to four numeric characters NOTE The security code can only be changed after the m...

Page 38: ...and together for more than 1 second during normal operation mode to enter unsecured mode The calibration security code entry is shown on the display Figure 1 4 Calibration security code entry display 2 Key in your security code Press or on to move the cursor to the left or to the right Press or to increment or decrement the digit NOTE If you forget your security code see To reset the calibration s...

Page 39: ...t security code is entered an error code will appear briefly on the secondary display after which the multimeter will prompt you for the security code again 4 Exit the unsecured mode by pressing and together to secure the multimeter again See Figure 1 5 for the multimeter display NOTE Key in the default factory security code if you have reset the security code or if you are unsecuring your multime...

Page 40: ...Multimeter display when unsecuring the multimeter Press and for more than 1 second Hz Save Esc Shift Press or on for left or right Press or to increment or decrement the digit Hold Vsense Press to save your input Hz Save Fail After appearing briefly After appearing briefly Pass ...

Page 41: ...security code is shown on the display If you are changing the calibration security code for the first time or if you have reset the security code to its factory default is shown on the display 2 Set your new calibration security code by entering any four digits ranging from to Press or on to move the cursor to the left or to the right Press or to increment or decrement the digit 3 Press to save th...

Page 42: ...de Figure 1 6 Multimeter display when changing the calibration security code Press for more than 1 second Esc Shift Press or on for left or right Press or to increment or decrement the digit Hold Vsense Press to save your input Hz Save After appearing briefly Pass ...

Page 43: ...C mode and will be shown on the secondary and primary display 4 Enter the last four digits of the multimeter s serial number Press or on to move the cursor to the left or to the right Press or to increment or decrement the digit 5 Press to confirm your entry If the four digits entered are correct the display will show The calibration security code is now set to its factory default code If the inco...

Page 44: ...ure 1 7 Multimeter display when resetting the calibration security code to its factory default Press for more than 1 second Esc Shift Press or on for left or right Press or to increment or decrement the digit Hold Vsense Press to save your input Hz Save Fail After appearing briefly After appearing briefly Pass ...

Page 45: ...ait for 3 to 5 seconds before you begin the calibration after connecting the test leads Calibrate the DC mV prior to ambient temperature calibration Do not calibrate the DC mV immediately after a high current measurement or calibration Turn on the multimeter for at least 1 hour with the Type K thermocouple connected before proceeding with the ambient temperature adjustment by connecting specified ...

Page 46: ... V 10 kHz 0 9 to 1 1 x Reference value 60 00 V 3 00 V 55 Hz Reference value 30 00 V 55 Hz 0 9 to 1 1 x Reference value 30 00 V 1 kHz 0 9 to 1 1 x Reference value 30 00 V 10 kHz 0 9 to 1 1 x Reference value 600 00 V 30 00 V 55 Hz Reference value 300 00 V 55 Hz 0 9 to 1 1 x Reference value 300 00 V 1 kHz 0 9 to 1 1 x Reference value 300 00 V 10 kHz 0 9 to 1 1 x Reference value 1000 00 V 30 00 V 55 H...

Page 47: ...e value Short SHort Short V Ω COM terminals AC mV 60 00 mV 3 00 mV 55 Hz Reference value 30 00 mV 55 Hz 0 9 to 1 1 x Reference value 60 00 mV 55 Hz 0 9 to 1 1 x Reference value 30 00 mV 1 kHz 0 9 to 1 1 x Reference value 30 00 mV 10 kHz 0 9 to 1 1 x Reference value 600 00 mV 30 00 mV 55 Hz Reference value 300 00 mV 55 Hz 0 9 to 1 1 x Reference value 600 00 mV 55 Hz 0 9 to 1 1 x Reference value 300...

Page 48: ... to 1 1 x Reference value 600 00 kΩ 300 00 kΩ 0 9 to 1 1 x Reference value 60 00 kΩ 30 00 kΩ 0 9 to 1 1 x Reference value 6 00 kΩ 3 00 kΩ 0 9 to 1 1 x Reference value 600 00 Ω 300 00 Ω 0 9 to 1 1 x Reference value 60 00 Ω 30 00 Ω 0 9 to 1 1 x Reference value Short SHort Short V Ω COM terminals Table 1 19 Adjustment input values continued Test function Step Reference value Valid reference input ...

Page 49: ...F 0 10 mF 0 9 to 1 1 x Reference value 1 00 mF 0 9 to 1 1 x Reference value 10 00 mF 1 00 mF 0 9 to 1 1 x Reference value 10 00 mF 0 9 to 1 1 x Reference value 100 00mF 1 00 mF 0 9 to 1 1 x Reference value 10 00 mF 0 9 to 1 1 x Reference value Diode Short SHort Zero reference point 3 10 V 2 00 V 0 9 to 1 1 x Reference value Short SHort Short V Ω COM terminals DC μA mA Open oPEn Open terminals 60 0...

Page 50: ...Reference value 10 00 A 55 Hz 0 8 to 1 2 x Reference value a The sensor for the Vsense non contact voltage calibration is located on the top center of multimeter Place the top area as close as possible to the standard signal source without any loads b Set the 5520A to internal reference Prior to performing adjustment connect one end of the K type thermocouple with miniature TC connector on both en...

Page 51: ...ess to start the calibration If the calibration is successful will be displayed 7 Set the source output to 30 V 55 Hz 8 Press to start the calibration process 9 Set the calibrator to standby mode when the calibration is done See Figure 1 8 for the multimeter display NOTE Ensure the multimeter is REMOVED from the holster prior to performing this calibration Keep the multimeter away from fluorescent...

Page 52: ...Procedures 52 Keysight U1281A U1282A Service Guide Figure 1 8 Multimeter display for Vsense non contact voltage calibration Press to start calibration Hz Save Fail After appearing briefly Pass After appearing briefly ...

Page 53: ... 5 Check whether the right standard value has been applied or see Calibration Error Codes on page 57 6 Repeat step 4 and step 5 until calibration is complete for all calibration values 7 Press or rotate the rotary switch to select other values to be calibrated Repeat step 6 and step 8 NOTE Review the Test Considerations on page 26 and Adjustment considerations on page 45 before beginning the adjus...

Page 54: ...ultimeter will return to normal measurement mode and is secured 9 Verify the adjustments using the Performance Verification Tests on page 27 Exiting the adjustment mode 1 Remove all the shorting plugs and connectors from the multimeter 2 Record the new calibration count 3 Press and together to exit calibration mode 4 Cycle the multimeter s power The multimeter will then be secured Hz Save Esc Shif...

Page 55: ...hen you receive your multimeter take note of the calibration counts as a record for maintenance The calibration count increases by one for each calibration point from up to the maximum Once the calibration count reaches the calibration count will reset to The calibration count can be read after the multimeter has been unsecured The calibration count cannot be programmed or reset It is an independe...

Page 56: ...1 Calibration Procedures 56 Keysight U1281A U1282A Service Guide Figure 1 9 Multimeter display for calibration count Press for more than 1 second Hold ...

Page 57: ... code Description 200 E200 Calibration error calibration mode is secured 002 E002 Calibration error secure code invalid 003 E003 Calibration error serial number code invalid 004 E004 Calibration error calibration aborted 005 E005 Calibration error value out of range 006 E006 Calibration error signal measurement out of range 007 E007 Calibration error frequency out of range 008 E008 EEPROM write fa...

Page 58: ...1 Calibration Procedures 58 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 59: ...ooting 60 Fuse Replacement 61 Returning the Multimeter for Service 64 Replaceable Parts 65 Types of Service Available 66 Obtaining Repair Service Worldwide 67 This chapter will help you troubleshoot a failing multimeter It also describes how to obtain repair services and lists the replaceable assemblies ...

Page 60: ...is not covered in this manual should only be performed by qualified personnel Table 2 1 Operating checklist Malfunction Identification No display when powered ON using the rotary switch Verify the condition of the batteries and replace batteries as necessary No beeper tone Verify that the beeper is enabled in the multimeter s setup mode Failed on current measurement Verify the condition of the fus...

Page 61: ... mm 440 mA 1000 V 30 kA fast acting fuse Fuse 1 The terminal is protected by a 10 x 38 mm 11 A 1000 V 30 kA fast acting fuse Fuse 2 If you are certain that the fuse is faulty replace it with one of the same size and rating NOTE No recalibration is required after replacing the fuse mA A CAUTION Before you proceed with the fuse replacement remove all cable connections to the terminals and ensure tha...

Page 62: ...Guide 1 Remove the orange rubber holster Pull a top corner to stretch the orange rubber holster off the multimeter 2 Loosen and remove the two screws with a suitable Phillips screwdriver as shown on the right 3 Lift and remove the fuse cover as shown on the left ...

Page 63: ...and removing it out of the fuse bracket 6 Replace a new fuse of the same size and rating at the center of the fuse holder 7 Ensure that the inner rubber cover is positioned properly 8 Close the inner rubber cover of the fuse compartment 9 Place the fuse cover back in its original position and tighten the screws 10 Finally fit the orange rubber holster back on the multimeter Fuse 1 for terminal mA ...

Page 64: ...imeter model number Multimeter serial number Description of the service required or failure indications 2 Remove all accessories from the multimeter Do not include accessories unless they are associated with the failure symptoms 3 Place the multimeter in its original container with appropriate packaging material for shipping If the original shipping container is not available place your unit in a ...

Page 65: ...lude a brief description of each part with its corresponding Keysight part number To order replaceable parts You can order replaceable parts from Keysight using the Keysight part number Note that not all parts listed are available as field replaceable parts To order replaceable parts from Keysight do the following 1 Contact your nearest Keysight Sales Office or Service Center 2 Identify the parts ...

Page 66: ...arranty After your warranty expires Keysight offers repair services at competitive prices Extended service contracts Many Keysight products are available with optional service contracts that extend the covered period after the standard warranty expires If you have such a service contract and your multimeter fails during the covered period Keysight will repair or replace it in accordance with the c...

Page 67: ...d or replaced and can provide warranty or repair cost information where applicable To obtain warranty service or technical support information you can contact Keysight using our Web link www keysight com find assist Or contact your Keysight representative Before shipping your multimeter request the Keysight Service Center to provide shipping instructions including what components to ship Keysight ...

Page 68: ...2 Service and Maintenance 68 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 69: ...ct to change without notice Always refer to the English version at the Keysight website for the latest revision Keysight Technologies 2016 2017 Edition 4 June 30 2017 Printed in Malaysia U1281 90002 U1281 90002 www keysight com ...

Reviews: