1
Calibration Procedures
44
Keysight U1241C/U1242C Service Guide
Valid adjustment input values
Adjustment can be accomplished using the following input values.
Table 1-15
Adjustment input values
Test function
Step
Reference value
Valid reference input
AC V
1000.0 mV
1000.0 mV (55 Hz)
0.9 to 1.1 x Reference value
1000.0 mV (1 kHz)
0.9 to 1.1 x Reference value
100.0 mV (55 Hz)
0.9 to 1.1 x Reference value
1000.0 mV (55 Hz)
0.9 to 1.1 x Reference value
1000.0 mV (10 kHz)
0.9 to 1.1 x Reference value
10.000 V
10.000 V (55 Hz)
0.9 to 1.1 x Reference value
10.000 V (1 kHz)
0.9 to 1.1 x Reference value
1.000 V (55 Hz)
0.9 to 1.1 x Reference value
10.000 V (55 Hz)
0.9 to 1.1 x Reference value
10.000 V (10 kHz)
0.9 to 1.1 x Reference value
100.00 V
10.00 V (55 Hz)
0.9 to 1.1 x Reference value
100.00 V (55 Hz)
0.9 to 1.1 x Reference value
100.00 V (1 kHz)
0.9 to 1.1 x Reference value
100.00 V (10 kHz)
0.9 to 1.1 x Reference value
1000.0 V
100.0 V (55 Hz)
0.9 to 1.1 x Reference value
1000.0 V (55 Hz)
0.9 to 1.1 x Reference value
1000.0 V (1 kHz)
0.9 to 1.1 x Reference value
1000.0 V (10 kHz)
0.9 to 1.1 x Reference value
Vsense
[a]
Hi.10 V
10 V (55 Hz)
High sense
Lo.30 V
30 V (55 Hz)
Low sense