Performance Tests and Calibration
5
Keysight U1241AN User’s and Service Guide
75
Table 5-4
Calibration adjustments
Step
Test function
Cal range
Input reference value
Cal item
U1241AN
1
Turn the rotary switch to
position
Short
Dual banana plug with
copper wires short
between two terminals
SHrt
1000 mV
1 V
1000.0 mV
10 V
10 V
10.000 V
100 V
100 V
100.00 V
1000 V
1000 V
1000.0 V
2
Press
to go to
function
1000 mV
30 mV, 70 Hz
30.0 mV
1000 mV, 70 Hz
1000.0 mV
1000 mV, 10 kHz
1000.0 mV
10 V
0.5 V, 70 Hz
0.500 V
10 V, 70 Hz
10.000 V
10 V, 10 kHz
10.000 V
100 V
5 V, 70 Hz
5.00 V
100 V, 70 Hz
100.00 V
100 V, 2.5 kHz
100.00 V
1000 V
50 V, 70 Hz
50.0 V
1000 V, 70 Hz
1000.0 V
1000V, 1 kHz
1000.0 V
3
Turn the rotary switch to
position
Short
Dual banana plug with
copper wires short
between two terminals
SHrt
1 V
1 V
1.000 V