:ACQuire:SEGMented:INDex
Command Syntax
:ACQuire:SEGMented:INDex <index>
<index> ::= an integer from 1 to 1000 (w/4M memory) in NR1 format
The :ACQuire:SEGMented:INDex command sets the index into the memory
segments that have been acquired.
The segmented memory acquisition mode is enabled with the :ACQuire:MODE
command. The number of segments to acquire is set using the
:ACQuire:SEGMented:COUNt command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments that have been
acquired is returned by the :WAVeform:SEGMented:COUNt? query. The time tag of
the currently indexed memory segment is returned by the
:WAVeform:SEGMented:TTAG? query.
The maximum number of segments may be limited by the memory depth of your
oscilloscope. For example, an oscilloscope with 1M memory allows a maximum of
250 segments.
Query Syntax
:ACQuire:SEGMented:INDex?
The :ACQuire:SEGMented:INDex? query returns the current segmented memory
index setting.
Return Format
<index><NL>
<index> ::= an integer from 1 to 1000 (w/4M memory) in NR1 format
See Also
•
•
•
•
•
•
•
•
•
"Introduction to :ACQuire Commands"
Example Code
' Segmented memory commands example.
' -------------------------------------------------------------------
Option Explicit
Public myMgr As VisaComLib.ResourceManager
Public myScope As VisaComLib.FormattedIO488
Public varQueryResult As Variant
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...