![Keysight P9241 Scpi Programmer'S Manual Download Page 1106](http://html1.mh-extra.com/html/keysight/p9241/p9241_scpi-programmers-manual_19616821106.webp)
1106
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
33
Obsolete and Discontinued Commands
:FUNCtion:SOURce
Command Syntax
:FUNCtion:SOURce <value>
<value> ::= {CHANnel<n> | ADD | SUBTract | MULTiply}
<n> ::= 1 to (# analog channels) in NR1 format
The :FUNCtion:SOURce command is only used when an FFT (Fast Fourier
Transform), DIFF, or INT operation is selected (see the:FUNCtion:OPERation
command for more information about selecting an operation). The
:FUNCtion:SOURce command selects the source for function operations. Choose
CHANnel<n>, or ADD, SUBT, or MULT to specify the desired source for function
DIFF (differentiate), INTegrate, and FFT operations specified by the
:FUNCtion:OPERation command.
Query Syntax
:FUNCtion:SOURce?
The :FUNCtion:SOURce? query returns the current source for function operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | ADD | SUBT | MULT}
<n> ::= 1 to (# analog channels) in NR1 format
See Also
•
"Introduction to :FUNCtion<m> Commands"
•
NOTE
The :FUNCtion:SOURce command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :FUNCtion:SOURce1 command (see
) instead.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...