382
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
17
:FUNCtion<m> Commands
:FUNCtion<m>:SOURce1
Command Syntax
:FUNCtion<m>:SOURce1 <value>
<m> ::= 1 to (# math functions) in NR1 format
<value> ::= {CHANnel<n> | FUNCtion<c> | MATH<c> | WMEMory<r> | BUS<b>}
<n> ::= 1 to (# analog channels) in NR1 format
<c> ::= {1}, must be lower than <m>
<r> ::= 1 to (# ref waveforms) in NR1 format
<b> ::= {1 | 2}
The :FUNCtion<m>:SOURce1 command is used for any :FUNCtion<m>:OPERation
selection. This command selects the first source for the operator math functions or
the single source for the transform functions, filter functions, or visualization
functions.
The FUNCtion<c> or MATH<c> parameters are available for the transform
functions, filter functions, and the magnify visualization function (see
"Introduction to :FUNCtion<m> Commands"
on page 349) when <c> is lower than
<m>.
In other words, higher math functions can operate on lower math functions when
using operators other than the simple arithmetic operations (+, - , *, /). For
example, if :FUNCtion1:OPERation is a SUBTract of CHANnel1 and CHANnel2, the
:FUNCtion2:OPERation could be set up as a FFT operation on the FUNCtion1
source. These are called cascaded math functions.
To cascade math functions, select the lower math function using the
:FUNCtion<m>:SOURce1 command.
When cascading math functions, to get the most accurate results, be sure to
vertically scale lower math functions so that their waveforms take up the full
screen without being clipped.
The BUS<m> parameter is available for the bus charting visualization functions.
When :FUNCtion<m>:OPERation is TRENd, the :FUNCtion<m>:SOURce1
command reports error -221,"Settings conflict" because the TRENd function
operates on a measurement and not a source waveform.
Query Syntax
:FUNCtion<m>:SOURce1?
NOTE
Another shorthand notation for SOURce1 in this command/query (besides SOUR1) is SOUR.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...