40 Keysight N9923-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
S21 and S12 transmission measurements are used to measure the loss or gain
through a DUT over a specified frequency range. Both ends of the DUT must be
connected to the FieldFox. The FieldFox signal source is transmitted out one of
the test port connectors, through the DUT, and into the other test port
connector
How to measure S-parameters
1.
Press
Preset
then
Preset
2.
Press
Measure 1
then choose from the following:
—
S11
Reflection measurement at port 1.
—
S21
Forward 2-port transmission measurement.
—
S12
Reverse 2-port transmission measurement. May require an
option.
—
S22
Reflection measurement at port 2. May require an option.
OR
select a multi-trace configuration. Learn more on
.
3.
Press
Freq/Dist
then either
Start
and
Stop
or
Center
and
Span
to enter a
Frequency Range for the measurement.
4.
Press
BW 2
then
IFBW
to select the IF Bandwidth for the measurement.
Narrower bandwidths require more time to sweep, but lowers trace noise.
5.
Press
Sweep 3
then
Resolution
to select the number of data points for the
measurement. More data points require more time to sweep.
6.
Press
Cal 5
to calibrate the measurement. Learn more in
.
7.
All other settings can be made AFTER calibration without compromising
measurement accuracy.
Mixed-Mode S-Parameters
In NA Mode ONLY, mixed-mode S-parameters (also known as Balanced
measurements) are available with Option 212.
Because the FieldFox has only two test ports, only reflection measurements are
available. Connect the balanced input or output of your DUT to the FieldFox
ports 1 and 2
For highest accuracy, a Full 2-port calibration is required.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...