120 Keysight N9923-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
Interpolated Calibrations are only slightly less accurate than a calibration
performed at the measurement settings. Learn more about the relative
accuracy of FieldFox calibrations in
“Cal ON ? – Questionable Accuracy” on
.
When a calibration that you performed is being interpolated, an asterisk is
added to the Cal annotation. For example:
Cal ON U*
is shown on the screen
when the current Response or Mechanical cal is being interpolated. An
*
is
never added to a CalRdy.
Cal ON ? – Questionable Accuracy
When the Output Power, Interference Rejection, or IF BW (NA Mode ONLY)
setting is changed AFTER performing a calibration, a question mark is added to
the Cal annotation. For example:
Cal ON Q?
is shown on the screen when the
current QuickCal is being interpolated. A
?
is never added to CalRdy.
The resulting measurement accuracy depends on how much the setting has
changed. For highest accuracy, recalibrate using the new settings.
Compatible Mode Calibrations
The FieldFox can have only ONE calibration present for all modes. Because NA,
CAT, and VVM modes are very similar, a calibration that is performed in one
mode can also be applied in the other modes with the same type of
measurements (1-port or 2-port).
To apply a Cal that was performed in a different mode, press
Cal 5
then select
Cal ON
.
Save the Calibration
After performing any type of calibration, you can save the FieldFox settings
along with the calibration into a STATE (*.sta) file. These settings and
calibration can then be recalled as necessary. To learn how, see
CalReady Properties
There are several factory calibrations (CalReady) on every FieldFox. These can
be selected based on the type of DUT that you measure most often, and the
compromise that you prefer to make between measurement speed versus
measurement accuracy. Remember, CalReady was performed at the test ports.
Therefore, a CalReady calibration is most accurate when the DUT is directly
connected to the test ports. Learn more in
.
This setting does NOT survive Preset or Power ON/OFF.
— Press
Cal 5
then
More
— Then press
CalReady
to toggle between the following selections:
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...