7-24 Keysight N9923A Service Guide
Repair and Replacement Procedures
Removing and Replacing the Front Panel Group Assemblies
7-
Figure 7-15
Replacing the A2 LCD, the LCD Shield, and LCD Cables
Replacing the A3 Front Panel Interface Board (FPIB)
1.
Perform the steps under
“Pre-replacement Procedure” on page 7-21
.
2.
Remove the four screws (item 2) that secure the LCD shield and remove
the LCD shield. Discard the four screws, new ones are provided in the
. Be careful. The A2 LCD display is now loose
and will fall out of the front case if the case is tipped.
3.
Disconnect the RPG cable, W12, at the A3 FPIB connector as indicated.
4.
Disconnect the A2 LCD to FPIB cable, W113, at the A3 FPIB connector as
indicated.
The A3 front panel interface board contains a Lithium-ion battery. Refer to
“Battery Disposal” on page 1-8
for important instructions on disposing of
this printed circuit board with the battery attached. If necessary, the
battery can be removed from the printed circuit board and disposed of
separately.