Keysight N9038A MXE Service Guide 299
Front End Control
A15 Front End Control Assembly Troubleshooting
11.
If the level is not correct, measure the input signal level coming to A15
. It should be at approximately the
same level shown in
If the input signal is at the correct level, and the output is not, the A15
Enhanced Front End Control Assembly has a problem.
If the input signal is not at the correct level see
for instructions on how to determine the cause
of this problem.
12.
Reconnect W16 to A15J 900 as shown in
The 300 MHz IF signal routing for the A25 Wideband Analog IF output of
the A15 Enhanced Front End Control board can be seen in the solid heavy
blue line in
Figure 8-15
300 MHz IF Switching to A25 Wideband Analog IF
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...