298 Keysight N9038A MXE Service Guide
Front End Control
A15 Front End Control Assembly Troubleshooting
The following procedure will verify the 322.5 MHz IF output switching:
1.
Remove the instrument dress cover and internal cover. Refer to
18, “Assembly Replacement Procedures.”
for instructions on removing
these covers.
2.
Turn the instrument on and allow it to complete its boot up process.
3.
Switch the instrument into the Spectrum Analyzer mode by pressing
Mode
,
Spectrum Analyzer
.
4.
Turn the instrument auto alignment routine off by pressing
System
,
Alignments
,
Auto Align
,
Off
.
5.
Turn the 50 MHz Calibrator signal on by pressing
Input/Output
,
RF
Calibrator
,
50 MHz
.
6.
Tune the instrument to a frequency of 50 MHz with a span of 0 Hz by
pressing
FREQ Channel
,
Center Freq
,
50 MHz
and
SPAN X Scale
,
Zero
Span
.
7.
Set the input attenuation to 10 dB by pressing
AMPTD Y Scale
,
Attenuation
,
10 dB
.
8.
Disconnect W16 from A15 J900 as shown in
.
9.
Connect a spectrum analyzer to A15 J900.
10.
Verify that the amplitude of the 322.5 MHz signal at A15 J900 is equal to
that shown in
, +/- 3 dB.
Figure 8-14
A15 Enhanced Front End Control Assembly 322.5 MHz IF Output
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...