1-28
N67xx DUT Power Supply Installation
Updating the Configuration Files
The power supply connections must be defined in the system and standard config
files:
\I3070_ICT_ROOT\diagnostics\th1\config
\I3070_ICT_ROOT\standard\config
Use the
supplies
statement.
supplies <type> <#> asru channel <#>
•
The supplies type can be ps6751, ps6752, ps6746, or ps6773 depending on
which power supply was installed.
•
The first number or range of numbers gives the testhead supply number
(
1
through
24
). The numbering of the supplies in the
supplies
statement is
arbitrary. The software will accept any mapping of supply numbers to modules.
However, you cannot assign a power supply number more than once in a
testhead. If the customer has more than one system, make the new
supplies
statement compatible with the other system.
•
The second number or range of numbers gives the ASRU Card DUT channel.
This must match the physical connection of the supply to the ASRU card. You
cannot assign a channel number (1 through 6) more than once in a module.
The following table shows the default and alternate power supply mapping for each
module and the proper syntax for each type of power supply.
After the
config
files have been updated, compile both files.
If the new power supply has been placed on a wrong channel,
you will get a compiler error.
Summary of Contents for N6746A
Page 4: ...iv ...