20 Installation Note N5225-90107
Description of the Upgrade
Installation Procedure for the Upgrade
-
Figure 8
Location of Cable Clamps for W29 (N5245-20117)
— W18 (N5245-20049) A32 port 2 receiver coupler to W17
— W85 (N5245-20026) A30 port 3 receiver coupler to A39 port 3 source
attenuator
— W89 (N5245-20026) A31 port 4 receiver coupler to A40 port 4 source
attenuator
— W99 (N5245-20073) Port 3 RCVR C IN to A47 port 3 receiver attenuator
— W86 (N5245-20027) A39 port 3 source attenuator to front-panel Port 3
SOURCE OUT
— W25 (N5245-20116) A30 port 3 receiver coupler to front-panel REF 3
SOURCE OUT
— W100 (N5245-20066) A47 port 3 receiver attenuator to A28 mixer brick
(C)
— W44 (reuse) (N5245-20020) REF 3 RCVR R3 IN to A28 mixer brick (R3)
, install two cable ties (part number 1400-0249) to