 Operation

7

Operation

1

Connect the probe to the BNC input of the oscilloscope.

2

Select the desired volts/division range. (If you can set probe 

attenuation on the oscilloscope, set it to 1000:1)

3

Whenever possible, turn the high voltage source off before 

making any connections.

4

Connect the divider probe ground lead (alligator clip) to a good 

earth ground or reliable chassis ground.

5

Connect the probe to the voltage source.

6

Before turning on the high voltage source, make sure that no 

part of the person holding the probe is touching the device 

under test. Once this is certain, turn on the high voltage source.

7

Measure the voltage under test and observe the waveform on 

the oscilloscope. Remember the actual voltage is 1000 times 

greater than the oscilloscope waveform if the probe attenuation 

on the oscilloscope has been set to 1:1.

8

Turn off the high voltage source.

9

Disconnect the Keysight N2771B High Voltage Probe from the 

high voltage source BEFORE disconnecting the ground clip 

lead.

WARNING

Do not attempt to take measurements from sources where the 

chassis or return lead is not grounded.

WARNING

The ground lead is critical to the safe operation of the probe. Failure 

to make this connection when making high voltage measurements 

may result in personal injury or damage to the probe or oscilloscope. 

This connection must always be made BEFORE the probe tip comes 

in contact with the high voltage and must not be removed until the 

probe tip has been removed from the high voltage source.

WARNING

Do not connect the ground clip lead to the high voltage source for 

any reason.

WARNING

Do not make any floating measurements with the Keysight N2771B 

High Voltage probe.

Summary of Contents for N2771B

Page 1: ...Keysight N2771B 15 kV High Voltage Probe User s Guide...

Page 2: ...odify distribute or disclose the Software The EULA and the license set forth therein does not require or permit among other things that Keysight 1 Furnish technical information related to commercial c...

Page 3: ...and nominal input capacitance between 6 and 20 pF The N2771B is a 1000 1 divider which extends the voltage measurement capability to maximum 15 kV DC AC peak 10 kVrms AC or 30 kVpeak 20 ns pulse widt...

Page 4: ...ory I CAT I only Do not use the probe for measurements performed on circuits defined by any other measurement category or transient overvoltages of more than 1500V Refer to IEC Measurement Category De...

Page 5: ...nnecting the probe WARNING The probe body should be kept clean and free of any conductive contamination Refer to the section on cleaning WARNING The ground lead is critical to the safe operation of th...

Page 6: ...II is for measurements performed on circuits directly connected to the low voltage installation Example Household appliances portable tools and similar equipment Measurement Category III CAT III Measu...

Page 7: ...r than the oscilloscope waveform if the probe attenuation on the oscilloscope has been set to 1 1 8 Turn off the high voltage source 9 Disconnect the Keysight N2771B High Voltage Probe from the high v...

Page 8: ...8 kV DC or 5 kV AC rms Keysight recommends that an interval of 5 minutes be taken between measurements WARNING This high voltage probe is designed for use within Measurement Category I CAT I only Do...

Page 9: ...imately 10 volts amplitude 4 Adjust the oscilloscope time base for 20 s div 5 Use a trimmer tool to adjust the compensation capacitor for the flattest pulse top See the figure below for the location o...

Page 10: ...tio For input capacitance of 6 pF to 20 pF 1000 1 Bandwidth Designed For Use In 50 MHz 3 dB POLLUTION DEGREE 2 Input Resistance Operating Temperature 100 M when terminated by 1 M 0 C to 50 C Cable Len...

Page 11: ...Voltage Derating 11 Voltage Derating The following graph shows the voltage versus frequency derating curve for the Keysight N2771B High Voltage Probe Figure 2 Voltage Versus Frequency Derating Curve...

Page 12: ...12 Contacting Keysight Contacting Keysight Information on contacting Keysight Technologies can be found at www keysight com find contactus...

Reviews: