Step 5: Verify Operation of the M9703A Module
5. Select
mode.
6. Select
persistence mode.
7. Ensure that
is enabled by clicking on the tick box under the instrument
name.
8. Select
Channel Combination mode, and
9. Make the following configurations:
Input Full Scale : 2V
Offset : 0.000000 V
Coupling : DC 50
Nbr of Segments : 1
Nbr of Samples : 200
Sampling Frequency: 3200 MS/s
10. Open the
control window and make the following configuration :
Module: Module 1
Trigger Class: Edge
Trigger Source: Ch1 (INPUT 1)
Trigger Level : 0 mV
Trigger Coupling: DC
Trigger Slope: Positive
11. Select the
Acquisition mode and the Waveform Display should be as
shown below.
20