Step 5: Verify Operation of the M9703A Module
Keysight M9703A Startup Guide
19
Perform a Verification of the M9703A (optional)
Self-Test / Calibration
The M9703A is capable of performing both a self-test and calibration internally. This
function is automatically carried out when the SFP application is started, and if no
subsequent error message is displayed then the operation has been successful.
These operations may also be carried out at any time by the user by pressing the
or
buttons on the control panel of the SFP.
It may take several minutes to complete a calibration operation.
Requirements for Verification
The M9703A is verified by using it to trigger on and visualize a signal from a Function
Generator. The trigger must be stable and the signal frequency and amplitude must
correspond to that set on the generator.
Required Hardware
To verify that the module works requires an external signal source. Almost any sine
wave or function generator capable of generating a signal with a Peak-Peak
Amplitude of 1 V into 50 Ω (10 dBm) at a frequency of 10 MHz may be used.
Hardware
Description
RF Analog Signal Generator
e.g. Keysight N5181A
1x BNC - SMA cable
50 Ω Coaxial BNC(m) to SMA(m) cable (100 cm)
1x Keysight 1250-1476 Adaptor
Type N(m) to BNC(f) adaptor
Operational Verification Procedure
Do not exceed the maximum power level to the INPUT connector (0.5 W).
1. Launch the soft front panel (SFP). Confirm that there are no error messages dis-
played.
2. Configure the RF Generator to produce a Sine signal with a Frequency of
10MHz, an Amplitude of +8.0 dBm, and an Offset of 0 V.
3. Connect the RF Generator output to the In 1 connector, and turn on the output.
4. Select the waveform mode.