Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
39
The
Δ
Y value on the lower line of the display should be within the test limits of
(depending on the oscilloscope model).
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then
return here.
10
Disconnect the oscilloscope calibrator from the oscilloscope.
11
Repeat this procedure to check the dual cursor accuracy with the remaining
Volts/div setting values in
(depending on the oscilloscope
model).
12
Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise
may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and
shorting cap at the oscilloscope channel input to shunt the noise to ground. See
. If a BNC capacitor is not available, use an SMA blocking capacitor,
adapter, and cap. See
“Blocking capacitor and shorting cap
for details.
Figure 6
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor