Setting Up the Oscilloscope
1
Keysight Infiniium S-Series Oscilloscopes User’s Guide
17
Connecting Oscilloscope Probes
1
Attach the probe connector to the desired oscilloscope channel or trigger input
using the probe instructions.
2
Connect the probe to the circuit of interest using the browser or other probing
accessories.
3
Disconnect the probe.
.
The Keysight Infiniium S-Series oscilloscopes are not rated for Measurement
Category II, III, or IV.
Figure 3
S-Series oscilloscope probe connectors
CAUTION
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50
Ω
input impedance setting is ±5 V.
The maximum input voltage for the 1 M
Ω
input impedance setting is 30 Vrms or
±40 Vmax (DC+Vpeak)
Probing technology allows for testing of higher voltages; the included N2873A 10:1
probe supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage
allowed.
Push
for Vernier
Push
to Zero
DIGITAL D15-D0
1
2
3
4
All Inputs
1M
Ω
±40V MAX
50
5V MAX
1M
~
~ 14pF
Probe Comp
+
3
4
Digital channels input
4 analog input channels where
probes are connected
CAUTION
When measuring voltages over 30 V, use a 10:1 probe.
Summary of Contents for Infiniium DSOS054A
Page 1: ...Keysight Infiniium S Series Oscilloscopes User s Guide...
Page 9: ...Keysight Infiniium S Series Oscilloscopes User s Guide 9 Using the Demo Wizard 51 Index...
Page 10: ...10 Keysight Infiniium S Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight Infiniium S Series Oscilloscopes User s Guide 1 Setting Up the Oscilloscope...
Page 48: ...48 Keysight Infiniium S Series Oscilloscopes User s Guide 2 Using the Oscilloscope...
Page 52: ...52 Keysight Infiniium S Series Oscilloscopes User s Guide 3 Online Help and Other Information...