Using InfiniiMax III+ Series Probes
1
InfiniiMax III+ Series Probes User’s Guide
23
Figure 15
Example InfiniiMode Probe Circuit Connections
The following measurements can be made
without
reconnecting the probe tips.
These equations use the voltages shown in
.
Differential Measurement
Common Mode Measurement
Single-Ended Measurement with +Lead
Single-Ended Measurement with –Lead
NOTE
Soldering the ground wires is not required when making differential or single-ended (+ or –
leads) measurements.
V
+lead
V
lead
–
–
4V 1V
–
3V
=
=
V
+lead
V
lead
–
+
2
----------------------------
4V 1V
+
2
---------------
2.5V
=
=
V
+lead
gnd
–
4V 0V
–
4V
=
=
V
lead
–
gnd
–
1V 0V
–
1V
=
=