148
InfiniiMax III+ Series Probes User’s Guide
8
Performance Verification for N7000-Series Probes
Figure 107
Measuring the Differential Input Resistance of Solder-In Probe Head
3
Connect the probe amplifier to the oscilloscope or power supply so it is
powered.
4
Connect the ZIF or solder-in probe head to the probe amp and insert it into the
PV fixture as shown
.
5
Depress the pincher fingers on the PV fixture so they open and carefully insert
the tip wires under the pincher. Release the pincher once the tips are inserted.
6
, measure the DC input resistance between the banana
plugs. Since one tip wire is connected to the signal line and the other tip is
connected to the PV fixture ground, this is a measurement of the differential
input resistance. It should be 100 k
W
±2% (98 to 102 k
W
).
7
To measure the single-ended input resistance, measure the resistance between
the signal plug of the banana adapter and the probe amplifier ground, which
can be accessed as shown in
(through the vent window of the probe
amplifier).
Figure 108
Measuring the Single-Ended Input Resistance the Solder-In Probe Head
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...