6 List Sequence Measurements
Marker
Notes
The selected marker is remembered even when not in the Marker menu and is used if a Search is done or a
Band Function is turned on or for Signal Track or Continuous Peak.
Preset
Marker 1
State Saved
The number of the selected marker is saved in instrument state.
Initial S/W
Revision
A.05.00
Help Map ID
3153
Couple Markers
When this function is On, moving any marker causes an equal X Axis movement of every other marker
which is On. An “equal X Axis movement” means the difference is preserved between each marker’s X Axis
value (in the fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker
being moved (in the same fundamental x-axis units).
parameter_table_23.52941
52.94118
Key Path
Marker
Remote Command
:CALCulate:LSEQuencer:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:LSEQuencer:MARKer:COUPle[:STATe]?
Example
:CALC:LSEQ:MARK:COUP ON
Preset
Off, presets on Mode Preset and All Markers Off
State Saved
Saved in State
Initial S/W Revision
A.05.00
Help Map ID
35256
All Markers Off
Turns off all markers.
parameter_table_23.52941
52.94118
Key Path
Marker
Remote Command
:CALCulate:LSEQuencer:MARKer:AOFF
Example
CALC:LSEQ:MARK:AOFF
Couplings
sets the selected marker to 1.
Preset
n/a.
Initial S/W Revision
A.05.00
Help Map ID
35257
222
Sequence Analyzer User's & Programmer's Reference