Keysight Power Distribution Unit Installation Guide
General Installation Information
Step
A B C D E F
7
Position the first PDU by placing it against the PDU brackets and align the inner holes that are
closest to the column on the brackets to the mounting holes in the PDU. Secure the PDU using
two 10-32 x 0.5” machine screws from the PDU kit.
If you are installing a second PDU place the second PDU against the PDU brackets and align the
outer holes on the brackets with the mounting holes of the PDU. Secure using 10-32 x 0.5”
screws from the kit. See Figure 6.
Figure 6
Attaching the PDUs
Step
A
B
C
D
E
F
8 Attach one end of the PDU ground cable assembly to the PDU at the bottom of the PDU
using one #10 ext. tooth star washer and one 10-31 x 0.5” machine screw. Make sure
the washer is between the PDU and the connector of the cable assembly. Make sure that
the cable does not block the PDU receptacles and the cable is not pinched. See Figure
7.