3
Remote Interface Reference
System-Related Commands
96
E3632A User’s Guide
System-Related Commands
DISPlay {OFF|ON}
Turn the front- panel display off or on. When the display is
turned off, outputs are not sent to the display and all
annunciators are disabled except the
ERROR
annunciator.
The display state is automatically turned on when you
return to the local mode. Press
to return to the local
state from the remote interface.
DISPlay?
This query returns the front- panel display setting. Returns
0
(OFF) or
1
(ON).
DISPlay:TEXT <quoted string>
Display a message on the front panel. The power supply will
display up to 12 characters in a message; any additional
characters are truncated. Commas, periods, and semicolons
share a display space with the preceding character, and are
not considered individual characters.
DISPlay:TEXT?
This query returns the message sent to the front panel and
returns a quoted string.
DISPlay:TEXT:CLEar
Clear the message displayed on the front panel.
OUTPut {OFF|ON}
Enable or disable the outputs of the power supply. When the
output is disabled, the voltage value is 0 V and the current
value is 20 mA. At
*RST
, the output state is OFF.