Keysight E1458A 96-Channel Digital I/O Module Service Guide
41
Troubleshooting
Service
Checking for Heat Damage
Inspect the assembly for signs of abnormal internally generated heat such as
discolored printed circuit boards or components, damaged insulation, or
evidence of arcing.
Checking Switches/Jumpers
Verify that the interrupt priority jumpers are set correctly (factory set at level 1).
Verify that the logical address switch is set correctly (factory set at 144). Verify
that the pull-up enable switches are set correctly (factory set to disable). Verify
that the port/flag combine switches are set correctly (factory set to Disable on all
ports). Refer to the Keysight E1458A User’s Manual for procedures to set or
check these switches/jumpers.
Checking the Keysight E1458A PCA
Check the following:
–
Verify that fuse F101 is good (see the replacement procedure).
–
Check connectors P1, P2, J4A, J4B, J5A, J5B for damage.
–
Perform Test 2-1 in Chapter 2. Replace the module if the test fails.
Self-test Error Codes
Table 4-3 shows the self-test error codes for the module. The meaning of each
code is given in the right-hand column. If a self-test failure occurs, cycle power
and repeat the test. If the problem reoccurs, the device may need repair.
Table 4-3.
Table 4-3
Self-test Error Codes
Error
1
1
ss = card number (with leading zero deleted)
Description
+0
+ss01
+ss02
+ss03
+ss10
+ss11
+ss200+n
Self-test passes
Read register 00h fails
Read register 02h fails
Bad ID information in ID register
Interrupt expected but not received
Busy bit was not held ≈10.5 to 18.5 msec
Port interrupt at port n fails