6
CHANnel Subsystem
176
Keysight 8990B Programming Guide
CHANnel{2|3}:PROBe:ID?
This query returns the identification of the probe connected to the specified
channel.
Syntax
Example
Remarks
– This command is only applicable for channels 2 and 3. If a channel other than
channel 2 or 3 is specified, error 701,“Applicable to channel 2 and 3 only” will
occur.
– If channel 1 or 4 is not connected, error –241,“Hardware missing; Sensor is not
found in channel <channel>.” will occur.
:PROB
?
CHAN
2
3
:ID
CHAN3:PROB:ID?
Queries the channel 3 probe identification.