Specifications
2
Keysight
87421/2A Power Supply Operating and Service Manual
19
Environment Specifications
Keysight
Model Number
87421A and 87422A
Temperature
•
Operating
0°C to 55°C
•
Storage
-40°C to 70°C
Humidity
•
Operating
15% to 95% RH at 40°C
•
Non Operating Supersoak
90% RH at 65°C for 24 hours
•
Condensation
After stabilizing at 50% RH/25°C, product shall be subjected to 95% RH at 40°C.
Product shall suffer no damage and operate within specifications within 15min.
Shock
•
End Use Handling Shock
Half-sine shock, < 3ms duration, 155in/s
Δ
V, minimum 3 shocks each face
•
Transportation Shock (Type 1)
Trapezoid shock, 30g accelaration, 292in/s
Δ
V, minimum 3 shocks each face
Vibration
•
Operating
Random: 0.0001
/hz for 5Hz to 350Hz, -6dB/octave for 350Hz to 500Hz,
minimum 10min/axis for all 3 axes
•
Survival
Random: 0.015
/Hz from 5Hz to 100 Hz, -6dB/octave from 100Hz to 137Hz,
0.008
/Hz from 137Hz to 350Hz, -6dB/octave from 350Hz to 500Hz,
minimum 10min/axis for all 3 axes
Swept Sine: Resonant seach from 5Hz to 500Hz, 1octave/min sweep rate, 5min
resonant dwell at 4 resonances/axis
Altitude
•
Operating
4,600m
•
Storage
15,300m
ESD immunity
•
Direct Discharge
4kV
•
Air Discharge
8kV
g
2
g
2
g
2
Summary of Contents for 87421A
Page 1: ...Keysight 87421 2A Power Supply Operating and Service Manual ...
Page 2: ......
Page 3: ......
Page 10: ...8 Keysight 87421 2A Power Supply Operating and ...
Page 12: ...10 Keysight 87421 2A Power Supply Operating and Service Manual Contents ...
Page 16: ...14 Keysight 87421 2A Power Supply Operating and Service Manual 1 Introduction ...
Page 22: ...20 Keysight 87421 2A Power Supply Operating and Service Manual 2 Specifications ...
Page 26: ...24 Keysight 87421 2A Power Supply Operating and Service Manual 3 Operating and Service ...