Performance Tests
3
Keysight 86205A Operating and Service Manual
27
Function Test
The functional test confirms your bridge is operating correctly. The procedure checks the nominal
insertion loss and coupling of the bridge.
Recommended equipment
1
Connect the equipment as shown in
Figure 3-1
Equipment setup for a thru measurement calibration
2
To perform a thru measurement calibration, press the following keys on 8753.
PRESET
MEAS, S21
CAL, CALIBRATE MENU
RESPONSE, THRU, DONE:RESPONSE
3
Connect the equipment as shown in
Table 3-1
Recommended equipment for function test
Equipment
Recommended model
Network analyzer
8753C (or 8753A with 85046A)
S-parameter test set
85046A or 85047A
50
W
type-N (m) load
909F option 012
Type-N cable
part number 8120-4781
Adapters 7 mm to type-N (f)
11524A
KEYSIGHT 8753C
NETWORK
ANALYZER